Analysis of Digital Image Rejection (RFA) inDiagnosticRadiology Services after the Application of Computed Radiography (CR) to the Hospitals in Area Semarang City Gatot Murti Wibowo; Dwi Rochmayanti; M. Irwan Katili
Jurnal Riset Kesehatan Vol 2, No 1 (2013): Januari 2013
Publisher : Poltekkes Kemenkes Semarang

Show Abstract | Download Original | Original Source | Check in Google Scholar | Full PDF (8055.191 KB) | DOI: 10.31983/jrk.v2i1.151

Abstract

The purpose of this study was to determine the rejection of digital image analysis procedure CR softcopy hereinafter referred to as RFA, and describe the profile and characteristics of the proportion image rejection (rejection rates) based on the operational conditions of CR.Quantitative research was conducted with the observational approach. Data were collected by random sampling from 1181 digital image system CR-1137 type 1 at A hospital and the total number of CR type 2 digital image at B hospital on the status of post-image processing. Profiles and characteristics of the CR-system rejection rate was 9.15% while the type 1 for type 2 CR-system is 4.57%. Largest percentage of CR image rejection on both failitas dominated by chest radiographic anatomy of the organ which is 69.44% (A) and 48.08% (B). The findings reject rates of 15.87% due to poor performance of x-ray equipment and distribution teridenfitikasinya reject rates based on radiographer in A hospital deserves special attention. Though the radiographer and the employment rate figures reject individually in “ A” hospital not correlated (0.67 more than p-value).
The Effect Of Variation Milliampere-Seconds (mAs) To Entrance Skin Exposure And Radiographic Anatomy Image In Examination of Columna Vertebra Lumbal With Ap Projection Siti Daryati; Gatot Murti Wibowo
Jurnal LINK Vol 8, No 3 (2012): September 2012
Publisher : Pusat Penelitian dan Pengabdian kepada Masyarakat, Poltekkes Kemenkes Semarang

Show Abstract | Download Original | Original Source | Check in Google Scholar | Full PDF (2397.815 KB) | DOI: 10.31983/link.v8i3.315

Abstract

Quantitative Descriptive research with experimental opproach was done with x-ray machiine type UD150l-30E. The data of experiment was taken with made eight radiographs and the ESE measurement is conducted using TLD which layed on surface of abdomen phantom or at central point. Exposure was taken with eight times with eight variation of mAs. It is about 8,10,12,14,16,18,20 and 22 mAs with 80 kV. The result shown that there is a effetct of variation mAs to ESE and quality of radiographic anatomy image. Higher mAs result higher ESE and increase mAs about two interval reslut average value is 142 mR. Variation of mAs result variation of image quality. A good image quality with 19,1 cm part thickness resulted with use 18 mAs and 20 mAs with 80 kV. Optimum mAs with 80 kV in this research’s is 18 mAs.
Implementation of Compliance Testing Program on a Direct Digital Radiography (DDR) X-Ray System at the Radiography and Imaging Laboratory of the Health Polytechnic Ministry of Health Semarang Gatot Murti Wibowo; Rasyid Rasyid; Ahkmad Haris Sulistiyadi
Jurnal LINK Vol 10, No 2 (2014): Mei 2014
Publisher : Pusat Penelitian dan Pengabdian kepada Masyarakat, Poltekkes Kemenkes Semarang

Show Abstract | Download Original | Original Source | Check in Google Scholar | Full PDF (566.403 KB) | DOI: 10.31983/link.v10i2.263

Abstract

Technological evolution from conventional to digital in diagnostic imaging bring a change in the implementation of Quality Control (QC) of radiology. As a part of the Quality Assurance (QA), QC Radiology is implemented to determine the eligibility condition of the radiography equipment used for the radiological examinations. A Direct Digital Radiography (DDR) unit has been installed in Radiography Laboratory (Lab 3), School of Radiodiagnostic and Radiotherapy, the Health Polytechnic of Semarang since February 2013. However, routine performance tests of the equipment have never been performed. This study aimed to obtain baseline data by assessing the equipment parameters of the X-ray and the digital-DDR systems (illumination collimator; collimator accuracy; beam alignment; tube leakage; kVp accuracy; timer accuracy; mA linearity; reproducibility; HVL, and dark image; nonuniformity; noise; bad pixels; lag respectively). Accuracy and reproducibility of the DDR unit were measured following the procedure of Perka Bapeten No. 9-2011 (for X-ray System) and IEC 62220-1: 2003 (for digital system. Data were analyzed descriptively by QC-DR software. Results showed that X-rays system and digital system had overall good performance. Signal non-uniformity was found in the value of Local Signal to Noise Ratio (LSNRNU) and Global Signal to Noise Ratio (GSNRNU. Non-uniformity caused by the characteristics of the detector and the lack of correction in system performance. Signal correction software needed to be developed to reach signal uniformities. Spatial resolution and low contrast detectability testing also needed to be performed to ensure the ability of the DDR unit to produce image resolution and contrast picture of the object.
Analysis and Design Software Evaluation Assessment Clinical Practice (PKL / PKN) in the Web-Based Program D-III and Radiotherapy Techniques Radiodiagnostic Semarang Sri Mulyati; Gatot Murti Wibowo; Darmini Darmini
Jurnal LINK Vol 9, No 3 (2013): September 2013
Publisher : Pusat Penelitian dan Pengabdian kepada Masyarakat, Poltekkes Kemenkes Semarang

Show Abstract | Download Original | Original Source | Check in Google Scholar | Full PDF (3173.02 KB) | DOI: 10.31983/link.v9i3.303

Abstract

Has done research for the analysis and design of software evaluation clinical practice assessment (PKL / PKN) WEB based on the Diploma Program Radiodiagnostic and Radiotherapy Techniques Semarang. Needs analysis has been conducted for designing software evaluation clinical practice assessment (PKL / PKN) Web-based programming using PHP and MySQL. Analyses were performed with several stages of system analysis, analysis of the weaknesses of the old system based PIECES, a new system needs analysis, feasibility analysis and design of software systems sistem.Desain done with methods for object-oriented programming (Object Oriented Programming) created using the Unified Modeling Language (UML) that is using the use case, the database and interface design. With the analysis and design of the software, will be developed (development) and monitoring and evaluation in the next stage of research.
Quantitative Measurement of The Knee Joint Cartilage Signal Intensity (SI) When T2map Sequence Applied to Define a Biomarker for Early Detection of Osteoarthritis (OA) Based on Bmi Groups Sugiyanto Sugiyanto; Gatot Murti Wibowo; Heru Trisikwanto
Jurnal Riset Kesehatan Vol 3, No 1 (2014): Januari 2014
Publisher : Poltekkes Kemenkes Semarang

Show Abstract | Download Original | Original Source | Check in Google Scholar | Full PDF (724.994 KB) | DOI: 10.31983/jrk.v3i1.232

Abstract

This quantitative study was an experimental design. A total of 18 healthy male and female volunteers, with differnt BMI involved in the study by consecutive random sampling. An expert Radiologist evaluated all the 38 T2map images being studied. The arbitrary SI data were measured using one millimeter ROI of the machine measurement tools and the values of T2map were based on a fitted exponential decay calculation ?generated from the graphs developed by the Window excel. The mean differences amongst T2map values on the sagital plan femoro-tibial cartilage and coronal plan femoro-tibial (medial and lateral) were statistically tested using One-way Anova, and all the T2map images also graded by the radiologist following the OA the grading levels as published Internationally. In conclusion, an increased TE on the application of multi-spin echo sequence T2map, causing the calculated T2map values v? aried not only in the sagital area of t?he knee joint cartilage of the femoral and tibial but also in the coronal femoro-tibial (medial and lateral) for both the male and female healthy volunteer subjects, and different BMI category (under-weight, normal-weight, verweight). The values ?T2map within and between BMI group and among different gender showed statistically no difference (p-value more than 0.05). The averaged T2map and its  tandard deviations have been used to define t?he T2map biomarkers (baseline T? 2map values) in each BMJ group subjects and between gender for both of the knee joint sagital and coronal planes despite the 1.5T MRI modality is not facilitated to commercial qMRI supporting software.