Angga Yosainto Bequet
Poltekkes Kemenkes Semarang

Published : 12 Documents Claim Missing Document
Claim Missing Document
Check
Articles

Found 12 Documents
Search

ANILISIS IMAGE QUALITY CT SCAN THORAX DENGAN VARIASI LUNG WINDOW KERNEL PADA MSCT SIEMENS SOMATOM EMOTION 6 IMAGE QUALITY ANALYSIS CT SCAN THORAX LUNG WINDOWS WITH KERNEL VARIATIONS USING MSCT SIEMENS SOMATOM EMOTION 6 Angga Yosainto Bequet; Yeti Kartikasari; Sri Mulyati; Susi Tri Isnoviasih
Jurnal Imejing Diagnostik (JImeD) Vol 5, No 2: July 2019
Publisher : Poltekkes Kemenkes Semarang

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.31983/jimed.v5i2.4431

Abstract

Background: To be able to show the Lung anatomy in detail in CT Scan Thorax lung window then required a sharp image quality. One way to improve image sharpness is to use kernel settings. The purpose of research is to know the difference of the quality of anatomical image with kernel variation CT Scan Thorax Lung Window and know the proper kernel selection to produce CT scan of thorax lung windows the bestMethods: Type of research is an experiment. The study was conducted using a CT image scan of thorax on axial lung window slices in cases of lung tumors in the kernel B50s, B60s, B70s, B80s, and U90s. The research was conducted by measuring the CT image quality of thorax scan on the windows lung by way of questionnaire assessment to 5 radiologist as the observer to CT Scan image to determine the level of clarity of anatomical criteria.Results: The results showed that there were significant differences in the quality of anatomical imagery with kernel variation of B50s, B60s, B70s, B80s, U90s with p-value 0.05. CT images can thorax lung windows that have the best anatomical image quality of the samples taken are shown by the use of the kernel U90s.Conclusions: There are significant differences in the quality of anatomical imagery with kernel variation of B50s, B60s, B70s, B80s, U90s. The kernel that generates image detail for CT thorax lung windows is the U90s kernel
Efektifitas Penambahan Source to Image Distance (SID) terhadap Penurunan Dosis Radiasi pada Pemeriksaan Radiografi Cranium Angga Yosainto Bequet; Panji Wibowo Nurcahyo; Akhmad Haris Sulistiyadi
Jurnal Imejing Diagnostik (JImeD) Vol 8, No 1: JANUARY 2022
Publisher : Poltekkes Kemenkes Semarang

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.31983/jimed.v8i1.8043

Abstract

Background: Cranial radiographs are routinely made in 2 projections, namely Antero Posterior (AP) and Lateral with a minimum distance of 100 cm from the radiation source to the image receptor. In addition, it has been shown that increasing the SID from 40 to 48 inches reduces the skin dose even when the requirement for an increase in mAs is considered. A 44-inch or 48-inch SID is recommended where departmental equipment and protocols allow. This study aims to determine the effectiveness of the addition of SID to decrease the radiation dose on cranium radiographs.Methods: This type of research is a pre-experimental quantitative. The study was conducted at the radiology laboratory of Diploma Three Program of Radiology Purwokerto, Semarang Health Politechnic of Ministry of Health. The research subject is a Radiographic Xray Cranial Phantom. At the time of exposure, the radiation dose was measured and the resulting image was measured CNR value.Results: The results of radiation dose measurements at each SID setting for AP cranium radiographs are shown in the table above. setting SID 100 cm produces a radiation dose of 5.58 mSv, SID 110 cm produces a radiation dose of 5.16 mSv, SID 120 cm produces a radiation dose of 5.02 mSv, SID 130 cm produces a radiation dose of 4.84 mSv, SID 140 cm produces a radiation dose of 4.64 mSv and SID 150 cm produces a radiation dose of 4.36 mSv. There are differences in radiation dose values between SID settings of 100 cm, 110 cm, 120 cm, 130 cm, 140 cm and 150 cm with a statistical test p-value 0.001. The results of statistical tests on the Contrast to Noise Ratio value on changes in the SID value of the AP cranium radiography examination showed no significant difference. The p-value of the statistical test is 0.274 which means there is no difference in the CNR value between the SID settings of 100 cm, 110 cm, 120 cm, 130 cm, 140 cm and 150 cm.Conclusions: An increase in SID causes a decrease in radiation dose. The increase in SID did not cause a significant difference to the quality of the Contrast to Noise Ratio on the AP cranium radiograph image.
Nilai Contrast to Noise Ratio (CNR) Radiograf Thorax PA antara menggunakan Grid dengan tanpa Menggunakan Grid Angga Yosainto Bequet; Luthfi Rusyadi; Fatimah Fatimah
Jurnal Imejing Diagnostik (JImeD) Vol 6, No 2: JULY 2020
Publisher : Poltekkes Kemenkes Semarang

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.31983/jimed.v6i2.5653

Abstract

Background: In the Thoracic Radiography Technique, applying the use of a grid to control scattered radiation almost never occurs, because the organ that is radiologically examined is classified as soft tissue dominance. Consideration of using grid is very important when the thickness of the thoracic organ exceeds 10 cm, thus potentially increasing scattering radiation which will affect the value of the Contrast to Noise Ratio (CNR). The purpose of this study is to examine the difference between the use of grid and non-grid for the chest radiography techniques in the context of CNR values.Methods: In the study, a hundred healthy patient uderwent the chest posteroanterior (PA) radiographic examination, a half of the total patients was examined with grid whereas the rest without the grid embedded. All the resulted images were analyzed by means of the pixel value measurements at the specific organs of interest (costea and pulmonary tissues) in inside and outside of the lung organs, using 1,5 mm of ROI  from Dicom software. The statistical analysis of the CNR values was performed by comparing the results between the groups.Results: There is a significant difference in the contrast values on the PA radiographs between those using the Grid and without using the Grid (p-value 0.001). The average contrast value on the chest radiograph with Grid is 2283.60, while without Grid is 1878.58. In the Noise values, it also deems significance (p-value = 0.001). The average of the noise values employing the Grid in the technique is 25.32, whereas without Grid is 17.84. In the Contrast to Noise Ratio (CNR), the diference seems to be significant (p-value 0.001). The average CNR radiograph of the PA chest using the Grid is 100.79, while the non-grid is 125.62.Conclusions: The application of the grid in radiography technique gives significant difference in the image results (contrast, noise, and CNR), when compare to the radiographic techniques without the grid. Using the grid in the Technique improve the qPA-thoracic quality of the chest images.
Perbandingan Nilai Index Exposure pada Radiografi Thorax dengan Perubahan Kombinasi Nilai mA dan s Berbeda pada Nilai mAs yang Sama Angga Yosainto Bequet; Susi Tri Isnoviasih
Jurnal Imejing Diagnostik (JImeD) Vol 8, No 2: JULY 2022
Publisher : Poltekkes Kemenkes Semarang

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.31983/jimed.v8i2.9324

Abstract

Background: In CR, there is an indicator called Exposure Index (EI) which displays how much exposure is used for a particular examination. The Exposure Index depends on the exposure factor, the total area of the detector exposed to radiation, and the object being exposed (Bontrager Lampignano, 2014). One of the exposure factors that is often regulated by the radiographer is mAs. Milliampere seconds (mAs) determine the number of X-rays produced. The purpose of this study was to determine the differences in the Exposure Index Values on Chest Radiography with Changes in the Combination of Different mA and s Values at the Same mAs Value.Methods: This research is a type of quantitative research with a pre-experimental design. The object of this research is the thorax phantom. Data were collected by exposing thorax the chest phantom, then recording the resulting Exposure Index value. mA and s value variations that are set are 25mA x 320 ms, 160mA x 50 ms and 400mA x 20ms. Data analysis was carried out using statistical tests.Results: From the statistical test results it is known that there is no significant difference in the Exposure Index Value on Thorax Radiography between the 25mA x 320 ms combination group, the 160mA x 50 ms combination and the 400mA x 20ms combination with a p-value 0.05 (p-value = 0.151) . The reason for the absence of a difference in the Exposure Index value is the variation in the combination of different mA and s values for the same mAs because with the same mAs, the quantity / intensity value of the X-ray radiation produced is also the same or there is no increase or decrease. With the same x-ray radiation intensity, the radiation received by the Imaging Plate is also the same so that the value of the Exposure index is also the same.Conclusions: There is no significant difference in the Exposure Index Value on Chest Radiography with Changes in the Combination of Different mA and s Values at the Same mAs Value
Kualitas Informasi Anatomi Radiograf pada Pemeriksaan Sternum dengan Variasi Penyudutan Arah Sinar Rizki Saputra; Angga Yosainto Bequet
Jurnal Imejing Diagnostik (JImeD) Vol 9, No 1: JANUARY 2023
Publisher : Poltekkes Kemenkes Semarang

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.31983/jimed.v9i1.9626

Abstract

Background: The sternum's radiographic examination involves the use of Right Anterior Oblique (RAO) and Lateral projections. However, if a patient cannot undergo RAO projection, Left Posterior Oblique (LPO) projection in the supine position and adjusting the central ray angle are alternatives. On another note, deviations from perpendicular alignment among the central ray, image receptor, and object can lead to distortions in shape, impacting image quality. Hence, a study is conducted to investigate the impact of beam angle variations on the quality of anatomical information in sternum radiographs to identify the optimal angle that yields the highest quality anatomical information in sternum radiographs. Methods: This research was quantitative research employing an experimental methodology. Data collection involved employing a single-body phantom, which underwent three exposures at each of the specified angle variations Results: Based on the results of the Kruskal Wallis test, there were significant differences in the quality of radiographic anatomic information (p value 0.05). The angle with the most optimal quality of radiographic anatomical information was an angle of 25° to the left with a mean rank value of 11.00. Conclusions: There was a significant difference in the quality of radiographic anatomic information (p value 0.05). The angle with the most optimal quality of radiographic anatomic information was the angle with the highest mean rank value, which was an angle of 25° to the left. 
Nilai Dosis Radiasi pada Pemeriksaan Radiografi Abdomen AP Susi Tri Isnoviasih; Asri Indah Aryani; Angga Yosainto Bequet
Jurnal Imejing Diagnostik (JImeD) Vol 10, No 2 (2024): JULI 2024
Publisher : Poltekkes Kemenkes Semarang

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.31983/jimed.v10i2.10813

Abstract

Background: Abdominal radiography examination is one of the examinations that is routinely carried out on the abdominal area to diagnose abnormalities in the abdominal area using radiation. In the stomach area there are gonads which are organs that are sensitive to radiation. The X-rays that come out of the X-ray tube are polyenergy, that is, they are composed of low energy x-ray, medium energy x-ray and high energy x-rays. Low energy x-rays have no benefit on the radiographic image information produced, but only increase the radiation dose on the skin surface. The additional x-ray filter is able to prevent low-energy x-rays from hitting the patient. The aim of this study is to determine the difference in the dose of radiation during the abdominal examination between the use of an additional filter and the use of an additional filter without an additional filter.Methods: The research was carried out at the Radiology Unit of the Prof. Dr. Margono Soekarjo Regional Hospital, Purwokerto and the Purwokerto Radiology Study Laboratory, Poltekkes Semarang with a sample of 50 samples with variations in the use of filters including 0 mm, 1 mm Al + 0.1 mm Cu, and 1 mm Al + 0.2 mm Cu and 2 mm Al. The data measured is the radiation dose On the surface of the Gonad organ's skin.Results: The results of measuring the average dose to the skin surface of the gonad organ are 180.26 mSv on a 0 mm Al filter, 83.32 mSv on a 1 mm Al + 0.1 mm Cu filter, 58.32 mSv on a 1 mm Al + 0 filter, 2 mm Cu and 119.49 mSv on a 2 mm Al filter. The greatest reduction in radiation dose occurred when using an additional filter of 1 mm Al + 0.2 mm Cu, amounting to 68% of that without using an additional filter. The decrease in the use of other filters was 54% when using an additional filter of 1 mm Al + 0.1 mm Cu and 34% when using an additional filter of 2 mm Al. The statistical tests showed there was a difference in the radiation dose during abdominal examination between using an additional filter and without using an additional radiation filter with a significance value of 0.000 (p-value 0.05).Conclusions: There is a difference in radiation dose during abdominal examination between using an additional filter and without using an additional filter with a significance value of 0.000 (p-value 0.05).
Nilai Entrance Skin Dose Tiroid antara menggunakan Thyroid Shield dan tanpa Thyroid Shield pada Pemeriksaan CT Scan Kepala Angga Yosainto Bequet; Panji Wibowo Nurcahyo; Susi Tri Isnoviasih
Jurnal Imejing Diagnostik (JImeD) Vol 10, No 1: JANUARY 2024
Publisher : Poltekkes Kemenkes Semarang

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.31983/jimed.v10i1.10816

Abstract

Background: CT scans have a much greater dose than conventional radiographic examinations. The thyroid organ is an organ that is sensitive to radiation which is located near the head, precisely in the inferior part of the head so that when a CT scan of the head is carried out, the thyroid organ will be exposed to radiation. The aim of this study was to determine the difference in thyroid skin dose entry value between using a thyroid shield and without a thyroid shield on a head CT scan.Methods: This study is a type of quantitative research with a pre-experimental design. The object of research is Entrance Skin Dose (ESE) of ThyroidOrgan on non-contrast head CT Scan examination. The research location is at the Radiology Department at the Regional General Hospital Prof. Dr. Margono Soekarjo Purwokerto. Sample data number is 50 people (25 samples of non-contrast head CT scans without thyroid shield, and 25 samples of non-contrast head CT scans with thyroid shield. Data was obtained by measuring the radiation dose using a dosimeter placed over the thyroid area during a non-contrast CT scan between using a thyroid shield and without using a thyroid shield. Data analysis was performed using statistical tests Results: The results of measuring the radiation dose to the samples showed that the average radiation dose to the skin surface of the thyroid organ on a non-contrast CT scan without using a thoracic shield was 884.96 µSv and using a thyroid shield was 428.64 µSv. This shows that the average radiation dose from the skin surface of the thyroid organ decreased by 456.32 µSv (51.56%).Conclusions: There is a significant difference in the amount of radiation dose to the surface of the skin of the thyroid organ in non-contrast head CT scans between using a thyroid shield and without using a shield with a dose reduction of 51.6% of the radiation dose without using a thyroid shield.
Perbedaan Nilai Dosis Radiasi Permukaan Kulit Gonad pada Radiografi Abdomen Antero Posterior pada Nilai mA dan s Berbeda dengan Nilai mAs yang Sama Isnoviasih, Susi Tri; Bequet, Angga Yosainto; Nurcahyo, Panji Wibowo; Puspita, Mega Indah; Putri, Nabila Pramitya; Mufida, Widya
Jurnal Imejing Diagnostik (JImeD) Vol 11, No 1 (2025): JANUARY 2025
Publisher : Poltekkes Kemenkes Semarang

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.31983/jimed.v1i1.12308

Abstract

Background: Abdominal radiography examination is one of the examinations that is often carried out and contains radiosensitive organs, namely the gonads. Milli Ampere Second (mAs) is one of the parameters regulated in making radiographs besides kV. mAs is the product of mA and s and there are often variations in the use of mA and s values to produce the same mAs value. This study's objective was to identifythe difference in radiation dose values for the surface of the gonad skin on anteroposterior abdominal radiographs with changes in different combinations of mA and s values.sMethods: This is pre-experimental quantitative research. The research object is body phantom. Data were collected by exposing abdomen radiography the body phantom and measured the amount of radiation dose to the skin surface of the gonad. mA and s setting : 160 mA 125 ms; 200 mA 100 ms; dan 250 mA 80 ms. Data analysis was carried out using statistical tests.Results: The average radiation dose to the skin surface of the gonad organs in the combination is 160 mA 125 ms; 200 mA 100 ms; and 250 mA 80 ms, namely 883.27 µGy; 883.09 µGy; and 883.72 µGy. Statistical test results show there is no significant difference between the combination of 160 mA 125 ms; 200 mA 100 ms; and 250 mA 80 ms, namely 883.27 µGy; 883.09 µGy with a significance of 0.748 (p-value 0.05). mAs determine the intensity of X-rays; the number of X-rays produced is proportional to milliamperes. As long as the mAs value remains constant, the amount of radiation intensity produced is the same, and the radiation dose is sameConclusions: There is no significant difference in the value of radiation dose to the skin surface of the gonad area on abdominal radiography with different combinations of mA and s values at the same mAs value
Kualitas Informasi Anatomi Radiograf pada Pemeriksaan Sternum dengan Variasi Penyudutan Arah Sinar Saputra, Rizki; Bequet, Angga Yosainto
Jurnal Imejing Diagnostik (JImeD) Vol. 9 No. 1: JANUARY 2023
Publisher : Poltekkes Kemenkes Semarang

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.31983/jimed.v9i1.9626

Abstract

Background: The sternum's radiographic examination involves the use of Right Anterior Oblique (RAO) and Lateral projections. However, if a patient cannot undergo RAO projection, Left Posterior Oblique (LPO) projection in the supine position and adjusting the central ray angle are alternatives. On another note, deviations from perpendicular alignment among the central ray, image receptor, and object can lead to distortions in shape, impacting image quality. Hence, a study is conducted to investigate the impact of beam angle variations on the quality of anatomical information in sternum radiographs to identify the optimal angle that yields the highest quality anatomical information in sternum radiographs. Methods: This research was quantitative research employing an experimental methodology. Data collection involved employing a single-body phantom, which underwent three exposures at each of the specified angle variations Results: Based on the results of the Kruskal Wallis test, there were significant differences in the quality of radiographic anatomic information (p value < 0.05). The angle with the most optimal quality of radiographic anatomical information was an angle of 25° to the left with a mean rank value of 11.00. Conclusions: There was a significant difference in the quality of radiographic anatomic information (p value < 0.05). The angle with the most optimal quality of radiographic anatomic information was the angle with the highest mean rank value, which was an angle of 25° to the left. 
Nilai Dosis Radiasi pada Pemeriksaan Radiografi Abdomen AP Isnoviasih, Susi Tri; Aryani, Asri Indah; Bequet, Angga Yosainto
Jurnal Imejing Diagnostik (JImeD) Vol. 10 No. 2 (2024): JULI 2024
Publisher : Poltekkes Kemenkes Semarang

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.31983/jimed.v10i2.10813

Abstract

Background: Abdominal radiography examination is one of the examinations that is routinely carried out on the abdominal area to diagnose abnormalities in the abdominal area using radiation. In the stomach area there are gonads which are organs that are sensitive to radiation. The X-rays that come out of the X-ray tube are polyenergy, that is, they are composed of low energy x-ray, medium energy x-ray and high energy x-rays. Low energy x-rays have no benefit on the radiographic image information produced, but only increase the radiation dose on the skin surface. The additional x-ray filter is able to prevent low-energy x-rays from hitting the patient. The aim of this study is to determine the difference in the dose of radiation during the abdominal examination between the use of an additional filter and the use of an additional filter without an additional filter.Methods: The research was carried out at the Radiology Unit of the Prof. Dr. Margono Soekarjo Regional Hospital, Purwokerto and the Purwokerto Radiology Study Laboratory, Poltekkes Semarang with a sample of 50 samples with variations in the use of filters including 0 mm, 1 mm Al + 0.1 mm Cu, and 1 mm Al + 0.2 mm Cu and 2 mm Al. The data measured is the radiation dose On the surface of the Gonad organ's skin.Results: The results of measuring the average dose to the skin surface of the gonad organ are 180.26 mSv on a 0 mm Al filter, 83.32 mSv on a 1 mm Al + 0.1 mm Cu filter, 58.32 mSv on a 1 mm Al + 0 filter, 2 mm Cu and 119.49 mSv on a 2 mm Al filter. The greatest reduction in radiation dose occurred when using an additional filter of 1 mm Al + 0.2 mm Cu, amounting to 68% of that without using an additional filter. The decrease in the use of other filters was 54% when using an additional filter of 1 mm Al + 0.1 mm Cu and 34% when using an additional filter of 2 mm Al. The statistical tests showed there was a difference in the radiation dose during abdominal examination between using an additional filter and without using an additional radiation filter with a significance value of 0.000 (p-value < 0.05).Conclusions: There is a difference in radiation dose during abdominal examination between using an additional filter and without using an additional filter with a significance value of 0.000 (p-value < 0.05).