Performance Comparison of Dymind DH-76 and Sysmex Xn-1000 Automated Hematology Analyzers Pusparini Pusparini; Alvina Alvina
INDONESIAN JOURNAL OF CLINICAL PATHOLOGY AND MEDICAL LABORATORY Vol 28, No 3 (2022)
Publisher : Indonesian Association of Clinical Pathologist and Medical laboratory

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.24293/ijcpml.v28i3.1907

Abstract

An automatic hematology analyzer is an essential instrument for the modern laboratory. A new hematology analyzer must undergo comparability testing with a currently used hematology analyzer. This study aimed to compare the performance of the Dymind DH-76 and Sysmex XN-1000 hematology analyzers. This study involved 96 EDTA blood samples from patients aged 18 to 85 years old at a hospital in Jakarta. A complete blood count was performed on each blood sample using two above-mentioned analyzers Dymind DH-76 and Sysmex XN-1000. The results from both instruments were compared by using the Pearson correlation test and the Passing-Bablok regression analysis to determine the agreement in performance between the two instruments. Between the Dymind DH-76 and Sysmex XN-1000, there was a high degree of agreement and correlation concerning the investigated parameters with r > 0.900 and p<0.001 for the parameters RBC, HGB, WBC, HCT, MCV, MCH, and neutrophil, monocyte, and eosinophil counts. The MCHC parameter had the values of r=0.797; p=0.0001, while the lymphocytes parameter had r=0.734 and p=0.0001. Only the basophils parameter showed a different result between the two aforementioned analyzers, with r=-0.179 and p=0.102. The majority of complete blood count parameters showed an excellent correlation and a high degree of agreement between the two instruments. The Dymind DH-76 hematology analyzer meets international standards (National Committee for Clinical Laboratory Standards/NCCLS) and can be used for hematological assay in the laboratory.
Performance Comparison of Dymind DH-76 and Sysmex Xn-1000 Automated Hematology Analyzers Pusparini Pusparini; Alvina Alvina
INDONESIAN JOURNAL OF CLINICAL PATHOLOGY AND MEDICAL LABORATORY Vol. 28 No. 3 (2022)
Publisher : Indonesian Association of Clinical Pathologist and Medical laboratory

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.24293/ijcpml.v28i3.1907

Abstract

An automatic hematology analyzer is an essential instrument for the modern laboratory. A new hematology analyzer must undergo comparability testing with a currently used hematology analyzer. This study aimed to compare the performance of the Dymind DH-76 and Sysmex XN-1000 hematology analyzers. This study involved 96 EDTA blood samples from patients aged 18 to 85 years old at a hospital in Jakarta. A complete blood count was performed on each blood sample using two above-mentioned analyzers Dymind DH-76 and Sysmex XN-1000. The results from both instruments were compared by using the Pearson correlation test and the Passing-Bablok regression analysis to determine the agreement in performance between the two instruments. Between the Dymind DH-76 and Sysmex XN-1000, there was a high degree of agreement and correlation concerning the investigated parameters with r > 0.900 and p<0.001 for the parameters RBC, HGB, WBC, HCT, MCV, MCH, and neutrophil, monocyte, and eosinophil counts. The MCHC parameter had the values of r=0.797; p=0.0001, while the lymphocytes parameter had r=0.734 and p=0.0001. Only the basophils parameter showed a different result between the two aforementioned analyzers, with r=-0.179 and p=0.102. The majority of complete blood count parameters showed an excellent correlation and a high degree of agreement between the two instruments. The Dymind DH-76 hematology analyzer meets international standards (National Committee for Clinical Laboratory Standards/NCCLS) and can be used for hematological assay in the laboratory.
Extension Program on Knowledge About Balanced Nutrition and Health Examination for Reducing Non-Communicable Diseases in the Baduy Community Patricia Budihartanti Liman; Yenny Yenny; Alvina Alvina; Laurensius Christian Fernando; Verawati Sudarma; Kartini Kartini
Jurnal Abdimas Kartika Wijayakusuma Vol 7 No 2 (2026): Jurnal Abdimas Kartika Wijayakusuma
Publisher : LPPM Universitas Jenderal Achmad Yani

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.26874/jakw.v7i2.1333

Abstract

Kanekes Village is home to the Baduy indigenous community, which has limited access to educational and healthcare services and continues to strongly uphold its traditional customs and values. Local health center data indicate that hypertension and diabetes mellitus are among the ten most prevalent diseases in the area. Although some community members have adopted relatively healthy dietary habits, their understanding of the Balanced Nutrition Guidelines, which serve as the national reference for healthy eating practices in Indonesia, remains limited. This community service program aimed to improve public knowledge of balanced nutrition while conducting early health screening as a preventive measure against non-communicable diseases (NCDs). The program employed health education sessions, interactive discussions, anthropometric assessments, blood pressure measurements, and examinations of blood glucose, cholesterol, and uric acid levels. Knowledge improvement was evaluated using pre-test and post-test assessments. The activity involved medical students from the Faculty of Medicine, Universitas Trisakti, and a medical student from Universiti Sains Malaysia. The results demonstrated a 25% increase in participants’ knowledge following the educational intervention. In addition, health examinations identified cases of malnutrition and various metabolic disorders among participants. The program effectively enhanced community awareness of balanced nutrition and provided preliminary insights into the health status of the community. Continuous health education and regular health monitoring are recommended to support efforts to reduce morbidity and mortality associated with NCDs, particularly among the residents of Kanekes Village, Banten Province.