Dewantara Journal of Technology
Vol. 2 No. 1 (2021): Dewantara Journal of Technology Volume 2 No 1

Determination of Suitability Test on Digital Mammography Using Voltage Accuracy Parameter

Fitriani Fitriani (Hasanuddin University)
Muhammad Yunus (Gorontalo State University)
Muh Fachrul Latief (Gorontalo State University)
A Indra Wulan Sari Ramadani (Gorontalo State University)
Meilan Demulawa (Gorontalo State University)
Haris Munandar (Gorontalo State University)
Muh. Nur Akbar (Gorontalo State University)



Article Info

Publish Date
15 May 2021

Abstract

Determination of suitability test on digital mammography have been conducted from 2016-2018 at a hospital in Makassar City using voltage accuracy parameter. This measurement was used detector, and X-ray multimeter. Statistically, voltage accuracy parameter has correlation value, r = 0.866 (perfect correlation). While the p value obtained in voltage accuracy parameter is p < 0.05. There are significant differences in value from 2016-2018. The result of the suitability test for voltage accuracy parameter is still within the tolerance limit as recommended by the IAEA.

Copyrights © 2021






Journal Info

Abbrev

djtech

Publisher

Subject

Agriculture, Biological Sciences & Forestry Automotive Engineering Civil Engineering, Building, Construction & Architecture Computer Science & IT Electrical & Electronics Engineering

Description

Dewantara Journal of Technology (Dewantara J. Tech) adalah terbitan berkala yang dapat diakses secara terbuka dan didistribusikan secara nasional. Jurnal ini melingkupi berbagai topik dalam disiplin ilmu teknik, termasuk studi konseptual, studi teoritis dan eksperimental, instrumentasi, dan studi ...