Claim Missing Document
Check
Articles

Found 1 Documents
Search

Determination of Suitability Test on Digital Mammography Using Voltage Accuracy Parameter Fitriani Fitriani; Muhammad Yunus; Muh Fachrul Latief; A Indra Wulan Sari Ramadani; Meilan Demulawa; Haris Munandar; Muh. Nur Akbar
Dewantara Journal of Technology Vol. 2 No. 1 (2021): Dewantara Journal of Technology Volume 2 No 1
Publisher : Akademi Teknologi Industri Dewantara Palopo

Show Abstract | Download Original | Original Source | Check in Google Scholar

Abstract

Determination of suitability test on digital mammography have been conducted from 2016-2018 at a hospital in Makassar City using voltage accuracy parameter. This measurement was used detector, and X-ray multimeter. Statistically, voltage accuracy parameter has correlation value, r = 0.866 (perfect correlation). While the p value obtained in voltage accuracy parameter is p < 0.05. There are significant differences in value from 2016-2018. The result of the suitability test for voltage accuracy parameter is still within the tolerance limit as recommended by the IAEA.