Makara Journal of Science
Vol. 8, No. 1

Construction and Characterization of Conductometric Biosensor for Determination of the Diazinon Concentration

Prayoga, Indrajid (Unknown)
Mulyasuryani, Ani (Unknown)
Prasetyawan, Sasangka (Unknown)



Article Info

Publish Date
25 Mar 2014

Abstract

Excessive diazinon residue in vegetables can endanger human health. Therefore, a simple, fast, and accurate method is needed to detect residue. A conductometric biosensor is a good choice because it also offers high selectivity and sensitivity. The principle of detection of the conductometric biosensor is based on enzymatic hydrolysis of diazinon into O,O diethyl phosphorothiate,2-isopropyl-6-methylpyrimidin-4-ol, and H+ catalyzed by organophosphate hydrolase (OPH). The optimum amount of organophosphate hydrolase added to the screen-printed carbon electrode (SPCE) modified with BSA-glutaraldehyde is 118.5 µg, while the optimum pH is 8.5. This biosensor has a response time of 30 sec, a linear dynamic range of 0 to 1 ppm, sensitivity of 42.21 µS/ppm, and limit of detection of 0.19 ppm.

Copyrights © 2014






Journal Info

Abbrev

publication:science

Publisher

Subject

Description

Makara Journal of Science publishes original research or theoretical papers, notes, and minireviews on new knowledge and research or research applications on current issues in basic sciences, namely: Material Sciences (including: physics, biology, and chemistry); Biochemistry, Genetics, and ...