VALENSI
Jurnal Kimia VALENSI Volume 9, No. 1, May 2023

Box-Behnken Experimental Design for Electrochemical Aptasensor Optimization on Screen Printed Carbon Electrode/Silica-Ceria

Zakiyyah, Salma Nur (Unknown)
Eddy, Diana Rakhmawaty (Unknown)
Firdaus, Muhammad Lutfi (Unknown)
Subroto, Toto (Unknown)
Hartati, Yeni Wahyuni (Unknown)



Article Info

Publish Date
31 May 2023

Abstract

This study aims to optimize the epithelial sodium channel (ENaC) electrochemical aptasensor with the Box-Behnken experimental design. ENaC is a protein that plays a role in sodium ion transport in several epithelial tissues and is associated with hypertension. The ENaC protein aptamer is held in place in the electrochemical aptasensor by a modified screen-printed carbon electrode (SPCE) of silica-ceria composite (SiO2-CeO2). The unique structure of a silica matrix with high biocompatibility can form composites through a hydrothermal process. The Box-Behnken (BBD) experimental design is an efficient optimization method of factors that affect the experiment at three levels. The FTIR results of the silica-ceria composites were 549.35 cm-1 (Ce-O), 1095.3 cm-1 (Si-O-Si), and 491.28 cm-1 (Si-O). Meanwhile, SPCE/silica-ceria characterized by differential pulse voltammetry (DPV) showed an increase in peak current [Fe(CN)6]3-/4- from 3.190 μA to 9.073 μA. Three experimental factors, aptamer concentration, streptavidin incubation time, and aptamer incubation time, were optimized with BBD and obtained at 0.5 μg.mL-1, 30 minutes, and 1 hour. The optimum conditions observed resulted in a selective current response for ENaC protein detection. The optimization results can be applied to aptamer-based ENaC protein detection in samples.

Copyrights © 2023






Journal Info

Abbrev

valensi

Publisher

Subject

Chemistry

Description

Jurnal Kimia Valensi is a biannual and peer-reviewed open access journal published by Department of Chemistry, Faculty of Science and Technology UIN Syarif Hidayatullah Jakarta. This journal covering all aspect of ...