Baby Incubator Based on PID Control With Kangaroo Mode (Kangaroo Mode and Humidity) Nur Fildzah Hidayati; Endro Yulianto; Abd Kholiq
Journal of Electronics, Electromedical Engineering, and Medical Informatics Vol 1 No 2 (2019): October
Publisher : Department of Electromedical Engineering, POLTEKKES KEMENKES SURABAYA and IKATEMI

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/jeeemi.v1i2.3

Abstract

Baby Incubator is one of the electromedical equipment used to provide protection to premature babies or low birth weight. Baby care in the baby incubator causes separation of mother and baby. A new innovation in the care of premature babies who bring their babies and mothers closer is Kangaroo Method Management (PMK). Kangaroo method treatment aims to create emotional attachment to the baby, so that the mother will confidently take care. The research and manufacture of this module uses a method of treating baby incubator by adding a reading of the room temperature when the kangaroo mode runs at a setting of 34 ° C - 37 ° C with the effect of room temperature at 18 ° C, 24 ° C, and 30 ° C and setting the humidity control automatically by measuring at a temperature of 32 ° C, then measuring the results immediately. Sensors used to detect temperature are LM35 sensors and sensors used to detect humidity using DHT22. Based on the measurement results obtained uncertainty values ​​on the measurement of humidity 2.1, the measurement of temperature stability can be seen from the results of the graph which shows that the room temperature is not too influential at the temperature of the baby incubator. This study has shown the development of baby incubators to help mothers who have babies born prematurely so that they can have a good bond with how to treat skin to skin. This study has proven that its accuracy is appropriate to be used to stabilize the condition of babies born prematurely and help to treat PMK in some hospitals. In the future, this research can be made and used in small clinics in villages at low costs.
Syringe Pump With Nearly Empty Based Microcontroller Atmega328 Lely Erica Putri; Muhammad Ridha Mak'ruf; Abd Kholiq
Journal of Electronics, Electromedical Engineering, and Medical Informatics Vol 1 No 2 (2019): October
Publisher : Department of Electromedical Engineering, POLTEKKES KEMENKES SURABAYA and IKATEMI

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/jeeemi.v1i2.5

Abstract

Syringe Pump is a tool used to give liquid medicine or food liquid into the patient's body in a certain amount and within a certain period of time on a regular basis. The purpose of this study is to facilitate monitoring of fluid in the syringe so that the hose is not installed continuously when the liquid has run out. The circuit consists of an Atmega328 microcontroller, a motor driver, and an optocoupler sensor. Setting the syringe, volume and flowrate is done at the beginning. To insert liquid, the motor must be run by the way the settings have been done and press the start button. Tools need supply from PLN grids. Calibration is done using IDA 4 Plus. This tool is equipped with the addition of alarms nearly empty and the bolus button. From the measurements taken, at the 20 ml syringe the biggest error occurred at the 5 ml volume point of 0.4% and at the 50 ml syringe the biggest error occurred at the 5 ml volume point of 0.280%.This module can be used according to its function, because the% error is still below the ± 5% standard.
Infant Warmer Equipped with Digital Weight Scales Muhammad Jundi Al'Aziz; Bambang Guruh Irianto; Abd Kholiq
Jurnal Teknokes Vol 14 No 2 (2021): September
Publisher : Jurusan Teknik Elektromedik, POLTEKKES KEMENKES Surabaya, Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/teknokes.v14i2.4

Abstract

Scales Scales in the world of health are used to measure human body weight such as baby scales. Newborns are very important to be weighed because it is used as a measure of the baby's health indication ranging from 2.4 kg to 4.2 kg. The author makes a tool for this to make it easier for users to weigh with a 7 segment display on the Infant Warmer tool and external calibration. By using a loadcell sensor with a maximum capacity of 5 kg, the loadcell can detect the weight of the load where the voltage generated by the loadcell of 0.7 mV at a load of 1 kg is amplified to 0.62 V by the PSA circuit using the AD620 IC and then processed by Arudino UNO as a microcontroller. The weight results will be displayed on the 7Segment display located on the Infant Warmer tool. In the study, the measured load included a weight of 0 kg to a maximum weight of 5 kg. The measurement of the data results was carried out 5 times each by comparing the modules that had been made with the standard weight, namely (lead). The data from the measurement results of the research module shows that when the weight of the measurement at 1 kg has an error percentage of 0.08%. Measurements at a weight of 2 kg have an error percentage of 0.05%. Measurement of weight 3 kg has a presentation error of 0.01%. Measurements at a weight of 4 kg have a presentation error of 0.02%. And measurements at a weight of 5 kg have an error percentage of 0.04%. Then the data from the measurement results of the research module shows the largest error presentation of 0.08% at a weight of 1 kg. And the data from the measurement results of the research module shows the smallest error presentation of 0.01% at a weight of 3 kg. Making a research module in the form of a scale placed on an infant warmer can make it easier for the wearer.
A Low Cost Baby Incubator Design Equipped with Vital Sign Parameters Lamidi Lamidi; Abd Kholiq; Muslim Ali
Indonesian Journal of Electronics, Electromedical Engineering, and Medical Informatics Vol 3 No 2 (2021): May
Publisher : Department of electromedical engineering, Health Polytechnic of Surabaya, Ministry of Health Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/ijeeemi.v3i2.3

Abstract

A baby incubator is a heating device for premature babies so that the baby's temperature matches the temperature in the mother's womb. Premature babies are babies born in abnormal conditions where the baby's body size is below average or the gestational age has not reached 9 months. The aim of this study was to create a compact system at a low cost in an infant incubator. Apart from stabilizing the temperature, it can also calculate the premature baby's heart rate and oxygen saturation value. The contribution of the system being built can detect the early symptoms of congenital abnormalities so that abnormalities detection becomes more practical. The temperature stability in this system uses a threshold system, where the sensor readings are compared with the set point value. Meanwhile, the SpO2 system uses a finger sensor attached to the baby's finger to detect oxygen saturation and heart rate. The signal from the sensor is processed in a signal conditioning circuit consisting of an astable circuit, filter and amplifier, LPF, and demultiplexer. Based on the results of testing and measurement, the difference produced by this tool is only 0.13 - 0.182 ° C. The biggest error in the system is 0.517%. Measurement of skin temperature was carried out on 10 respondents with an average value of 34.825 ° C. so that the difference generated from this tool is only 0.175 ° C, the difference on the Humidity parameter is 0.21%. The average BPM reading of respondent 1 was 81 beats per minute and the average SpO2 value was 97%. The average BPM reading of respondent 2 was 83 beats per minute and the average SpO2 value was 98%. You can see that the module with the Arduino control system can work automatically and is able to adjust the stability of the baby incubator according to the settings. In addition, the BPM and SpO2 systems are within tolerance so they can be used for patient measurements