UJI KESESUAIAN CT NUMBER IN WATER PADA PESAWAT CT SCAN MERK PHILIP DI INSTALASI RADIOLOGI RSU. BMC PADANG : THE COMPATIBILITY TEST OF CT NUMBER IN WATER ON CT SCAN AIRCRAFT BRAND PHILIP IN RADIOLOGY INSTALLATION OF BMC PADANG GENERAL HOSPITAL Nansih, Livia Ade; Artitin, Cicilia
JURNAL TERAS KESEHATAN Vol 6 No 1 (2023): Jurnal Teras Kesehatan
Publisher : Lembaga Penelitian dan Pengabdian kepada Masyarakat (LPPM) Politeknik Al Islam Bandung

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.38215/jtkes.v6i1.105

Abstract

CT number is the value of the X-ray attenuation coefficient determined by the average energy of X-rays and the absorber atomic number is expressed by the attenuation coefficient. The role of the CT number on a CT scan machine to assess and differentiate abnormalities in human organs. Compliance tests are carried out to ensure that the CT-Scan aircraft meets radiation safety requirements and provides precise and accurate diagnostic information. This research is a quantitative research with an experimental approach, carried out at the Radiology Installation of RSU Bunda Padang in June 2022 for 5 days. The population of all CT number tests, samples at five measurement points with ROI in the center of the image, at 12 o'clock, 3 o'clock, 6 o'clock and 9 o'clock directions. Research data was taken by observing, experimenting and documenting. Data analysis was compared with PERKA BAPETEN Number 9 of 2011 concerning quality assurance of QA and QC of CT Scan aircraft. The results showed that the CT number value in the middle of the image on the first day was 4.83 HU, the second was 4.83 HU, the third was 4.14 HU, the fourth was 5.15 HU and the fifth was 5.16 HU while the average standard deviation on the first day was 3, 48 HU, the second day 3.42 HU, the third day 3.41 HU, the fourth day 3.43 HU and the fifth day 3.50 HU, the ROI uniformity results at the image and noise center show that it has exceeded the tolerance limits set by PERKA BAPETEN No. 9 of 2011.
Increasing Radiographers' Knowledge Through Socialization of Research Results Pocket Book in The Radiology Unit of Reksodiwiryo Hospital in Padang Nansih, Livia Ade; Sari, Oktavia Puspita; Chairunnisa, Chairunnisa; Mareta, Santa; Dewilza, Nerifa; Hanifah, Wahdini; Artitin, Cicilia; Yudha, Sagita
MSJ : Majority Science Journal Vol. 2 No. 3 (2024): MSJ-AUGUST
Publisher : PT. Hafasy Dwi Nawasena

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.61942/msj.v2i3.234

Abstract

Radiology is a medical science that focuses on imaging technology. with results in the form of images /images for. to establish a diagnosis by looking at parts of the human body using X-rays so as to be able to provide professional examination services. The results of research that have been conducted at Aulia Hospital regarding the selection of CT scanogram settings for head scans with clinical chepalgia using the upper limit of the vertex and the lower limit of the mandibular because the location of the abnormality in the patient is not known whether in the nerve or sinus, if the scanogram is made to the lower limit of the petrous pars it still cuts the cerebrum so that the sinuses are still cut. whereas in the Radiology installation of Reksodiwiryo Padang Hospital uses a scanogram of the lower limit of the cranii base and the upper limit of the vertex to minimize the radiation received by the patient. Based on this, socialization activities through a pocket book of research results were carried out at Reksodiwiryo Padang Hospital on August 10, 2023 starting at 08.00 - 12.00 WIB, with participants being radiographers in the Radiology installation of Reksodiwiryo Padang Hospital.  The methods used were lecture, discussion and question and answer. With communication that is two way traffic, there is a direct reciprocal relationship between the speaker and the participants or vice versa. This method applied can focus the attention of participants in training to develop thinking power and skills to express opinions in the form of exchanging information in solving a problem. The results of the socialization of the pocket book of research results on the effectiveness of CT scan management of the head with clinical chepalgia were 98% able to increase radiographers' knowledge in the selection of scanogram settings on CT scans of the head with chepalgia able to produce images that enforce dignosa
Lead apron leak test at the Padang Reksodiwiryo Army Hospital's radiology installation Rahmadianti, Yori; Yudha, Sagita; Dewilza, Nerifa; Nansih, Livia Ade; Ramadanti, Tasya
Journal Of Health Science (Jurnal Ilmu Kesehatan) Vol 8 No 1 (2023): JOURNAL OF HEALTH SCIENCE (JURNAL ILMU KESEHATAN) (IN PRESS)
Publisher : Fakultas Ilmu Kesehatan

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.24929/jik.v8i1.2496

Abstract

A lead apron is a lead apron designed to protect the body from radiation hazards. The lead apron test is carried out once a year to ensure that it can provide optimal protection when in use, it is necessary to test personal protective equipment. Based on observations, the storage and placement of tin aprons is not in accordance with the rules so that it can reduce its function as personal protective equipment. This study aims to determine whether there is a leak in the lead apron, and to find out whether the lead apron is still suitable for use as personal protective equipment (PPE). This type of quantitative research with experimental methods at the Radiology Installation of RST Dr. Reksodiwiryo Padang on March 31, 2022. The sample in this study was 2 lead aprons, each lead apron is divided into four quadrants, namely the upper right side, upper left side, lower right side and lower left side. Each quadrant was tested with a tube voltage of 70 Kv and a current of 16 mAs. Then made observations and compared with the theory of Lambert and Mc Keon. The results of the study found that there was a leak in the Apron A quadrant 3 with an area of ​​0.75 mm2 and quadrant 4 with an area of ​​0.67 mm2. On Apron B there is a leak in quadrant 4 with an area of ​​0.06 mm2. From the results of the study of lead apron leaks in quadrant 3 and quadrant 4, the quadrant area is a critical area which according to Lambert's theory should not exceed 15 mm2 so that the lead apron under study is still suitable for use as personal protective equipment (PPE).
UJI KEBOCORAN APRON MENGGUNAKAN PESAWAT SINAR-X FLUOROSKOPI DI RSUD M. NATSIR SOLOK TAHUN 2021: . Nansih, Livia Ade
JURNAL TERAS KESEHATAN Vol 5 No 1 (2022): Jurnal Teras Kesehatan
Publisher : Lembaga Penelitian dan Pengabdian kepada Masyarakat (LPPM) Politeknik Al Islam Bandung

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.38215/jtkes.v4i2.74

Abstract

The Apron is a lead apron to protect the body from radiation hazards, to reduce exposure to fluoroscopy radiation, and made of lead as a shield. Based on observations, the number of aprons was seven, purchased in 2011 and 2012. The apron was placed on a small table, and the wrong apron was found. Seven of the aprons, one of which was on the first apron, were suspected to have been damaged due to the wrong way of storing the apron and the researcher noticed kinks in the suspected apron. This type of research is quantitative, in the Radiology Unit of RSUD. Mohammad Natsir Solok from January 2021. Population 7 apron, sample 7 apron, data processing by measuring using tools available in fluoroscopy, data analysis compared with Lambert and McKeon (2001) theory. The results showed that the leak in the upper one apron was 51.54 mm2, the lower part 38.21 mm2 in the lower two apron 51.43 mm2, the third apron in the middle right front area was 21.12 mm2, in the lower right front area 1.664,87 mm2, at the rear center area 1,435.57 mm2, the top five rear area apron 6.97 mm2, at the rear center area 26.98 mm2 
Uji Kesesuaian CT Number In Water Pada Pesawat CT Scan Merk Philip Di Instalasi Radiologi RSU BMC Padang Nansih, Livia Ade; Artitin, Cicilia
JURNAL TERAS KESEHATAN Vol 6 No 1 (2023): Jurnal Teras Kesehatan
Publisher : Lembaga Penelitian dan Pengabdian kepada Masyarakat (LPPM) Politeknik Al Islam Bandung

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.38215/jtkes.v6i1.105

Abstract

CT number is the value of the X-ray attenuation coefficient determined by the average energy of X-rays and the absorber atomic number is expressed by the attenuation coefficient. The role of the CT number on a CT scan machine to assess and differentiate abnormalities in human organs. Compliance tests are carried out to ensure that the CT-Scan aircraft meets radiation safety requirements and provides precise and accurate diagnostic information. This research is a quantitative research with an experimental approach, carried out at the Radiology Installation of RSU Bunda Padang in June 2022 for 5 days. The population of all CT number tests, samples at five measurement points with ROI in the center of the image, at 12 o'clock, 3 o'clock, 6 o'clock and 9 o'clock directions. Research data was taken by observing, experimenting and documenting. Data analysis was compared with PERKA BAPETEN Number 9 of 2011 concerning the quality assurance of QA and QC of CT Scan aircraft.