Claim Missing Document
Check
Articles

Found 6 Documents
Search
Journal : Journal of Electronics, Electromedical Engineering, and Medical Informatics

Five Channel Temperature Calibrator Using Thermocouple Sensors Equipped With Data Storage Rizkiyatussani; Her Gumiwang Ariswati; Syaifudin
Journal of Electronics, Electromedical Engineering, and Medical Informatics Vol 1 No 1 (2019): July
Publisher : Department of Electromedical Engineering, POLTEKKES KEMENKES SURABAYA and IKATEMI

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/jeeemi.v1i1.1

Abstract

A temperature calibration device is a tool used to measure the accuracy of a temperature-related device such as a sterilisator. This temperature calibration device is needed when the temperature in the sterilisator is not linear. In this calibration tool the sensor used is a type-k thermocouple that is inserted into the media to be measured then the temperature results will be read. This tool is designed using pre-experimental methods with the type of after only design research. In this tool is equipped with storage on the micro sd card and also conversion mode to convert temperature results from Celsius to Rheamur, Farenheit and Kelvin. Temperature results will be displayed on a 4x20 LCD and processed using Arduino UNO. This module can be used in medical equipment calibration laboratories. After testing the thesis module with a comparison device from BPFK, the biggest error is obtained at 1% at 50 ° Celsius, 100 ° Celsius and 150 ° Celsius. The smallest percentage of error is 0% at 50 ° Celsius and 150 ° Celsius. It can be concluded that the tool "Temperature Calibrator (5 Channels) Using Thermocouple Equipped with Data Storage.
A Two-Mode Digital Pressure Meter Equipped With An Automatic Leak Test Using MPX5050gp And MPXv4115vc6u Sensors Fita Florensa Rooswita; Triana Rahmawati; Syaifudin Syaifudin
Journal of Electronics, Electromedical Engineering, and Medical Informatics Vol 2 No 2 (2020): July
Publisher : Department of Electromedical Engineering, POLTEKKES KEMENKES SURABAYA and IKATEMI

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/jeeemi.v2i2.7

Abstract

The calibration process aims to guarantee that a measurement device is to follow the established standards. The purpose of this study is to design an automatic leak test for digital pressure meter in which the function of this device is to measure pressure on the sphygmomanometer and suction pump or other devices that use pressure parameters for measurement. This study used the Arduino as a control system, and to process the analog data into digital. Signal conditioning, based on the amplifier circuit, was also applied for the MPX5050GP and MPXV4115VC6U sensor. The proposed design used a 4x20 liquid crystal display to show the parameters in this design ware also equipped with a selector of mmHg or kPa units. The result shows that the fluctuating resolution is 0.25 mmHg. In this design, an automatic leak feature was also equipped for the sphygmomanometer. The results obtained an average error of 7.3 mmHg for sphygmomanometers. On the other hand, the suction pumps have an accuracy of less than 1.5 kPa. From these results, it was concluded that this design could be used for the measurement of devices that use positive pressure and negative pressure
Oximeter and BPM on Smartwatch Device Using Mit-App Android with Abnormality Alarm Bedjo Utomo; Syaifudin Syaifudin; Endang Dian Setioningsih; Torib Hamzah; Parameswaran Parameswaran
Journal of Electronics, Electromedical Engineering, and Medical Informatics Vol 3 No 2 (2021): July
Publisher : Department of Electromedical Engineering, POLTEKKES KEMENKES SURABAYA and IKATEMI

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/jeeemi.v3i2.4

Abstract

Monitoring is an activity that is carried out continuously. Healthy condition is a parameter that is needed in life, one of the important parameters is the measurement of oxygen saturation in the blood and heart rate. The purpose of this research is to develop a Smartwatch SpO2 device and BPM sensor that is connected to WIFI using the Android Platform instead of using an LCD for parameter reading. This module design method uses the MAX30100 sensor to display the SpO2 and BPM values ​​displayed on the OLED. Data processing is carried out using ATMEGA 328P programming and then displayed in the Android-based Mit-app application. The results show the average error for the SPO2 value is 0.868 % and the standard deviation is 0.170 %, while the BPM value has an average error of 0.56 % and a standard deviation of 0.30%. From the results of the comparison data analysis, the largest error was 1.03% and the smallest was 0.62% for Spo2 ml/hour with an accuracy of 0.05 (0.57%) with a precision value of 0.08 at the selection speed of 50 ml/hour. From the results above, it can be concluded that the data can be displayed on OLED using the Mit-app Android application with an error rate accuracy of 0.57%. From the results of this research design, it is hoped that it can facilitate the diagnosis of the condition of patients and health nurses
Mechanical Fetal Simulator for Fetal Doppler Testing Arum Triwerdani; Syaifudin Syaifudin; Bedjo Utomo; Abdul Basit
Journal of Electronics, Electromedical Engineering, and Medical Informatics Vol 4 No 2 (2022): April
Publisher : Department of Electromedical Engineering, POLTEKKES KEMENKES SURABAYA and IKATEMI

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/jeeemi.v4i2.5

Abstract

The continuous use of fetal Doppler allows for discrepancies in values ​​that lead to misdiagnoses in patients. This study aims to determine the effect of sound source distance on the fetal simulator with the measurement point. The contribution of this research is that the mechanical fetal heart system has 4 distances so that later it can be analyzed whether there is an influence of the location of the sound source on the accuracy of measurements using a fetal simulator. To get the desired distance, a solenoid is used which ends with a pipe of 2 cm, 5 cm, 10 cm, and 50 cm respectively. The solenoid used in the fetal simulator functions as a producer of the fetal heart. There is a rotary switch that functions for solenoid selection, namely 2 cm, 5 cm, 10 cm and 50 cm solenoids. Data collection was carried out on each solenoid and by placing the Doppler probe perpendicular and tilted. On the solenoid with a distance of 50 cm all measurement results exceed the allowable tolerance limit. The results showed that the BPM value of the two Doppler brands did not have a significant difference in value. When measuring fetal Doppler, the largest error value was 2.67%. The results of this study can be used as a reference when conducting an examination
Sensor Accuracy Analysis on Incubator Analyzer to Measure Noise and Airflow Parameters Arrum Sekarwati; Syaifudin Syaifudin; Torib Hamzah; Shubhrojit Misra
Journal of Electronics, Electromedical Engineering, and Medical Informatics Vol 4 No 3 (2022): July
Publisher : Department of Electromedical Engineering, POLTEKKES KEMENKES SURABAYA and IKATEMI

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/jeeemi.v4i3.227

Abstract

Infant incubators are equipment to maintain a stable body temperature for premature babies. Premature babies need room conditioning that is close to conditions in the womb. Room conditioning is carried out in a baby incubator by providing a stable temperature, relative humidity, and measured air flow. This parameter must be controlled so as not to exceed the threshold that will harm the baby. Periodic calibration should be applied to the infant incubator to monitor its function. To ensure the availability of baby incubators according to service standards, it is necessary to conduct test (calibrate) using an incubator analyzer. The purpose of this study is to conduct further research on the incubator analyzer that focuses on discussing the accuracy of noise and airflow sensors with the gold standard. In this study, an experiment was carried out for the sensitivity level of several sensors that had been treated by giving treatment to sensors to choose sensors with good sensitivity to be assembled into one in the incubator analyzer module. The noise sensors (KY-037 and Analog Sound Sensor V2.2) were further compared with the values ​​on the sound level meter and the airflow sensor (D6F-V03A1) was compared with the anemometer. Sensors whose values ​​are close to the comparison values ​​were selected to be integrated into the incubator analyzer module. The incubator analyzer module used Arduino Mega2560 as a data processor and was equipped with an SD Card for the data storage. The built incubator analyzer module was also compared to the Fluke INCU II gold standard for data analysis. The results showed that the Analog Sound Sensor V2.2 had the highest error value (-4.6%) at 32°C and the D6F-V03A1 had the ability to measure sensitivity, where the results were more accurate than INCU II. Based on the error value of the noise sensor, the V2.2 sensor can be applied to measure noise in the baby incubator and the D6F-V03A1 airflow sensor produced an accuracy of up to 3 digits behind the comma which is more accurate than the standard module. The results of the INCU analyzer from this study can be used to calibrate the baby incubator, so that the certainty of the feasibility of the baby incubator is guaranteed. This research can be used as a reference for other researchers who will develop research on incubator analyzers in the future.
The Performance Analysis of the Infrared Photodiode Sensor to Infusion Set on Infusion Device Analyzer Machine Anisa Rahma Astuti; Syaifudin; Triana Rahmawati; Khongdet Phasinam
Journal of Electronics, Electromedical Engineering, and Medical Informatics Vol 5 No 1 (2023): January
Publisher : Department of Electromedical Engineering, POLTEKKES KEMENKES SURABAYA and IKATEMI

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/jeeemi.v5i1.274

Abstract

Infusion pumps and syringe pumps are devices used to administer liquid medicines to patients. The frequency of using the infusion pump and syringe pump in the long term will affect the accuracy of the tool. Accuracy is very important in dosing to patients in critical conditions who require intensive care to prevent fluid balance in the body. Therefore, periodic calibration of medical devices is required at least once a year. Calibration according to Permenkes No. 54 of 2015 is a calibration activity to determine the correctness of a tool. The purpose of this research is to make an Infusion Device Analyzer (IDA) with a TFT LCD displaying a graph of flowrate parameters. The method used is to analyze the flowrate value using an infrared photodiode sensor and can see the stability of the flowrate graph on a 7-inch TFT LCD from the use of 2 brands of syringes and an infusion set. The results obtained can be stored on the SD Card. The measurement results show that the error in the performance of the syringe and infusion pump read by the module on Channel 1 with the Terumo syringe is 0.15 (10 ml/h), 0.1 (50ml/h) and 0.03 (100ml/h). . On Channel 2 it is 0.02(10ml/hour), 0.03 (50ml/hour) and 0.04(100ml/hour). When using the B-Braun Channel 1 syringe, 0.25 (10ml/h), 009(50ml/h) and 0.08(100ml/h) are used. And on Channel 2 it is 0.62 (10ml/h), 0.15 (50ml/h), and 7.3 (100ml/h). When using the Terumo Channel 1 brand infusion set at 0.05 (10ml/h), 0.3(50ml/h), and 0.04(100ml/h). On Channel 2 it is 0.14(10ml/hour), 0.02 (50ml/hour) and 0.18 (100 ml/hour). When using the B-Braun Channel 1 Infusion Set, it is 0.07(10ml/h), 0.02(50 ml/h), and 0.03 (100ml/h). Then on Channel 2 0.07 (10ml/hour), 0.02(50 ml/hour), and 0.1(100ml/hour). The conclusion of this study is that the use of 2 different infusion sets greatly affects the readings, other than that other factors can also affect the readings including the position of the hose and the placement of sensors on each channel. From the manufacture of this tool, it is expected that users can be more efficient in using a 2-channel Infusion Device analyzer which can be run at the same time.