Digital Pressure Meter Equipped with Temperature and Humidity Abdul Cholid; Her Gumiwang Ariswati; Syaifudin Syaifudin
Indonesian Journal of Electronics, Electromedical Engineering, and Medical Informatics Vol 2 No 1 (2020): February
Publisher : Department of electromedical engineering, Health Polytechnic of Surabaya, Ministry of Health Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/ijeeemi.v2i1.1

Abstract

Calibration is a technical activity which consists of the determination, the determination of one or more properties or characteristics of a product, process or service in accordance with a special procedure has been set. The purpose of which is to ensure the calibration measurement results in accordance with national and international standards. The tools used for the calibration of pressure Digital Pressure Meter. This tool is used to measure the pressure and suction pump spygnomanometer or other devices that use parameters for measuring pressure. This module manufacturing system using Arduino system as a controller and as processing analog data into digital data of the sensor MPX5100GP and MPXV4115V using analog signal conditioning circuit and displayed on the LCD Touchscreen with 2 modes of measurement that is positive pressure and vacuum pressure with pressures ranging from 0-300 mmHg for positive pressure and 0 –(-400) mmHg to vacuum pressure. There are also DHT22 sensor, As a detector for temperature and humidity for use in the work method in the calibration process. Based on a stress test generated and using comparators Digital Pressure Meter 2 plus brand fluke, this tool has an error value of 0 to 0.58% and has a value increment or correction value of 0 - 3. It can be concluded that the DPM DUA MODE this deserves to be used.
Incubator Analyzer Using Bluetooth Android Display (Humidity & Air Flow) Vina Nadhirotul Azkiyak; Syaifudin Syaifudin; Dyah Titisari
Indonesian Journal of Electronics, Electromedical Engineering, and Medical Informatics Vol 1 No 2 (2019): November
Publisher : Department of electromedical engineering, Health Polytechnic of Surabaya, Ministry of Health Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/ijeeemi.v1i2.5

Abstract

Abstract— Incubator Analyzer is a calibrator used to calibrate incubator temperature, mattress temperature, noise, humidity and airflow so that conditions in the baby incubator environment remain stable and within normal limits. The purpose of this study is to develop a system for sending data to android and data storage. "Incubator Analyzer Using Bluetooth Appear Android" has four parameters for measuring temperature, noise, humidity, and water flow. Using the Atmega328 microcontroller as a data processor, and the output results will be displayed on the LCD display and equipped with data transmission via bluetooth HC-05 displayed to Android with data storage. The moisture parameter detects humidity quite well where the biggest error is obtained at 1.28% DHT-22, the Ultrasound Sensor HC SR-04 can detect Air Flow with an error of 311.66% as measured by a comparison device. Incubator Design This analyzer is made portable to calibrate baby incubator tools.
Temperature Calibrator with Thermocouple Based Microcontroller Aninda Zakia Febriyanti; Priyambada Cahya Nugraha; Syaifudin Syaifudin
Indonesian Journal of Electronics, Electromedical Engineering, and Medical Informatics Vol 2 No 1 (2020): February
Publisher : Department of electromedical engineering, Health Polytechnic of Surabaya, Ministry of Health Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar

Abstract

Calibration is very important to know the temperature uniformity inside the tool. One of them is at the temperature sterilizer, dry heating sterilizer which is sterilizing the device using a high heat oven. The purpose of this study is to develop a temperature Calibrator tool as a refinement of previously made tools by measuring temperatures more than one point in order to achieve a calibration process that complies with the standard. The working method of a temperature calibration device is that the sensor will detect the temperature which then enters the IC ATMEGA 328 that has been given the program and processed in such a way that the output will be displayed on the LCD 4x20 character in the form of temperature measurement of the device. Based on the results of comparative data between the module and the comparison tool "8 Channel Thermocouple Temperature Recorder" from the BPFK Surabaya, it has the largest difference of 40C and the percentage error (1,6%) and the smallest difference of 10C and the percentage error (0,16%).
Waterbath Calibrator with Nine Channels Sensor Mohammad Rofi'i; Syafudin Syaifudin; Dyah Titisari; Bedjo Utomo
Indonesian Journal of Electronics, Electromedical Engineering, and Medical Informatics Vol 1 No 1 (2019): August
Publisher : Department of electromedical engineering, Health Polytechnic of Surabaya, Ministry of Health Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar | Full PDF (495.44 KB) | DOI: 10.35882/ijeeemi.v1i1.1

Abstract

Water bath is a laboratory equipment that contains water or special liquid that can maintain the temperature under certain conditions during the specified time interval. For this reason, calibration is needed so that the temperature in the chamber waterbath is stable or not. calibration is carried out by comparing measuring instruments and measuring materials to be calibrated to traceable standards that are traceable to national and / or international standards. Based on the results of the identification of the problems mentioned above, the author makes a waterbath calibrator entitled "Waterbath Calibrator (9 channel)" which is very practical, and easy to operate. .This calibration tool uses a K type thermocouple sensor and also the output is displayed to the character LCD to make it easier for users to retrieve data, the reason for choosing a thermocouple sensor is because the error rate is +/- 1,1C while the LM35 is +/- 1,4C. The thermocouple temperature sensor can detect the chamber temperature quite well where the biggest error is obtained with a comparator of 2%, and the lowest error is 0%.
Design of Two Channel Infusion Pump Analyzer Using Photo Diode Detector Syaifudin Syaifudin; Muhammad Ridha Mak’ruf; Sari Luthfiyah; Sumber Sumber
Indonesian Journal of Electronics, Electromedical Engineering, and Medical Informatics Vol 3 No 2 (2021): May
Publisher : Department of electromedical engineering, Health Polytechnic of Surabaya, Ministry of Health Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/ijeeemi.v3i2.5

Abstract

In the medical world, patient safety is a top priority. The large number of workloads and the frequency of using the devices in the long run will affect the accuracy and accuracy of the tool. If the flow rate and volume of the syringe pump or infusion pump given to the patient are not controlled (overdose or the fluid flow rate is too high) it can cause hypertension, heart failure or pulmonary edema. Therefore, it is necessary to have a calibration, which is an application activity to determine the correctness of the designation of the measuring instrument or measuring material. The purpose of this research is to make a two channel infusion device analyzer using a photodiode sensor. The contribution of this research is that the system can display three calibration results in one measurement at the same setting and can calibrate 2 tools simultaneously. The design of the module is in the form of an infrared photodiode sensor for reading the flowrate value. This study uses an infrared photodiode sensor for channels 1 and 2 installed in the chamber. This study uses a flow rate formula that is applied to the water level system to obtain 3 calibration results. Infrared photodiode sensor will detect the presence of water flowing in the chamber from an infusion or syringe pump. Then the sensor output will be processed by STM32 and 3 calibration results will be displayed on the 20x4 LCD. This tool has an average error value on channel 1 of 3.50% and on channel 2 of 3.39%. It can be concluded that the whole system can work well, the placement and distance between the infrared photodiodes also affects the sensor readings
Analysis of Drop Sensor Accuracy in Central Infusion Peristaltic Monitoring Based on Computer Using Wireless Communication HC-11 Syaifudin Syaifudin; Triana Rahmawati; Siti Rohmatul Jannah; Sandeep Kumar Gupta; Ram Gopal
Indonesian Journal of Electronics, Electromedical Engineering, and Medical Informatics Vol 4 No 2 (2022): May
Publisher : Department of electromedical engineering, Health Polytechnic of Surabaya, Ministry of Health Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/ijeeemi.v4i2.2

Abstract

In some hospitals, the infusion is still done manually, where medical personnel observe the liquid droplets directly and then control the rate using a mechanical resistor (clamp). This method is certainly far from accurate. An infusion pump is a medical aid that functions to control and ensure the correct dose of infusion fluid given to patients treated. The purpose of this study was to analyze the accuracy of the infrared photodiode as a drop sensor based on the readings of the infusion pump monitoring system. This module consists of a photodiode infrared drop sensor module, a comparator circuit, a monostable circuit, a stepper motor, an L298N motor driver, and an ATmega328 microcontroller. The droplets were detected by an infrared photodiode sensor, then compared with a comparator and monostable circuit as an oscillator developer, and then the flow rate and residual volume readings were generated by the ATmega328 microcontroller. Next, this data has sent to the computer via the HC-11 wireless. The results of the flowrate module measurement show the highest error value of 3% at the 30 ml/hour setting and the lowest error value of 2.5% at the 60 ml/hour setting. Meanwhile, the results of the flow rate measurement using an infusion device analyzer obtained the highest error value of 4% at the setting of 30 ml/hour and 60 ml/hour, and the lowest error value of 0.8% at the setting of 100 ml/hour. Monitoring the infusion pump was designed centrally to facilitate the nurse's task in monitoring the infusion dose given to the patient accurately. Based on this research, the accuracy of the infrared sensor and photodiode is very good by looking at the existing error rate.
Comparative Analysis of Water and Oil Media on Temperature Stability in PID Control-Based Digital Thermometer Calibrator Mochammad Sofyan; Syaifudin Syaifudin; Andjar Pudji; Anggara Trisna Nugraha; Bedjo Utomo
Indonesian Journal of Electronics, Electromedical Engineering, and Medical Informatics Vol 5 No 2 (2023): May
Publisher : Department of electromedical engineering, Health Polytechnic of Surabaya, Ministry of Health Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/ijeeemi.v5i2.274

Abstract

Digital thermometers are measuring instruments needed to perform temperature measurement actions and must be calibrated periodically according to standard measurement methods. The purpose of developing this tool is to add PID control to the calibration media where PID control aims to regulate the stability of the temperature setting to be achieved. This is achieved by studying and evaluating the effect of temperature stability on the heater and LM35DZ temperature sensor. This research method uses the Arduino Nanosystem for data processing and PID system control. The LM35DZ temperature sensor on the heater is regulated by a 2 Channel SSR module using a PID system then the temperature generated by the heater will be read by the LM35DZ and displayed on the LCD. The results of this study, digital thermometer calibrator measurements have been successfully carried out by comparing 3 digital thermometers with different brands, namely Omron 343F, Omron 245, and ThermoOne. The difference in error values in oil media is 3-4% and in water media 2-4% with the value of time stability in water media for 3-3.3 minutes and in oil media for 1-2 hours. With this comparison of calibration media, it is hoped that it can help in measuring temperature with better and more effective results. find methods, results, conclusions.