Claim Missing Document
Check
Articles

Found 31 Documents
Search

Design and Build a Ventilator Tester with a Peak Inspiratory Flow Waveform Display as Validation using the F1031V Sensor Rusdi Pratiwo Hadi; Her Gumiwang Ariswati; Dyah Titisari; Syaifudin Syaifudin; Sari Luthfiyah; Bedjo Utomo; Liliek Soetjiatie
Jurnal Teknokes Vol 15 No 4 (2022): December
Publisher : Jurusan Teknik Elektromedik, POLTEKKES KEMENKES Surabaya, Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/teknokes.v15i4.472

Abstract

The ventilator is a supporter of respiratory needs which is very important for the patient so that there are several parameters that must be monitored specifically, such as the measurement of pressure and flow rate used in the ventilator system, the accuracy of which must be in accordance with the accuracy of the respirator. One of the important parameters to monitor is PIF (Peak Inspiratory Flow) which is the peak inspiratory flow rate given through the ventilator. PIF that is too high or too low can cause adverse effects on the patient. PIF monitoring can be seen through the PIF value and waveform on the PIF. Monitoring the waveform of the PIF will be very useful to improve the results of using the ventilator. The purpose of this research is to get the accuracy and precision of the sensor to display the waveform of the ventilator output. The procedure carried out is to use the F1031V sensor to detect the flow generated by the ventilator and then detect the PIF value and PIF waveform. From this research, the measurement of accuracy and precision of the F1031V sensor to detect PIF and generate a waveform graph is said to be good. This is because the highest error value is ±2.04% at the 20 LPM setting. While the value of the largest standard deviation at the 30 LPM setting is 1.517 and the greatest uncertainty value at the 30 LPM setting is 0.061. Then, the largest correction value is found in the setting of 20 LPM and 30 LPM, namely 0.4. PIF monitoring is carried out to maximize patient care and reduce the breakdown time on the ventilator.
The Performance Analysis of the Infrared Photodiode Sensor to Infusion Set on Infusion Device Analyzer Machine Anisa Rahma Astuti; Syaifudin; Triana Rahmawati; Khongdet Phasinam
Journal of Electronics, Electromedical Engineering, and Medical Informatics Vol 5 No 1 (2023): January
Publisher : Department of Electromedical Engineering, POLTEKKES KEMENKES SURABAYA and IKATEMI

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/jeeemi.v5i1.274

Abstract

Infusion pumps and syringe pumps are devices used to administer liquid medicines to patients. The frequency of using the infusion pump and syringe pump in the long term will affect the accuracy of the tool. Accuracy is very important in dosing to patients in critical conditions who require intensive care to prevent fluid balance in the body. Therefore, periodic calibration of medical devices is required at least once a year. Calibration according to Permenkes No. 54 of 2015 is a calibration activity to determine the correctness of a tool. The purpose of this research is to make an Infusion Device Analyzer (IDA) with a TFT LCD displaying a graph of flowrate parameters. The method used is to analyze the flowrate value using an infrared photodiode sensor and can see the stability of the flowrate graph on a 7-inch TFT LCD from the use of 2 brands of syringes and an infusion set. The results obtained can be stored on the SD Card. The measurement results show that the error in the performance of the syringe and infusion pump read by the module on Channel 1 with the Terumo syringe is 0.15 (10 ml/h), 0.1 (50ml/h) and 0.03 (100ml/h). . On Channel 2 it is 0.02(10ml/hour), 0.03 (50ml/hour) and 0.04(100ml/hour). When using the B-Braun Channel 1 syringe, 0.25 (10ml/h), 009(50ml/h) and 0.08(100ml/h) are used. And on Channel 2 it is 0.62 (10ml/h), 0.15 (50ml/h), and 7.3 (100ml/h). When using the Terumo Channel 1 brand infusion set at 0.05 (10ml/h), 0.3(50ml/h), and 0.04(100ml/h). On Channel 2 it is 0.14(10ml/hour), 0.02 (50ml/hour) and 0.18 (100 ml/hour). When using the B-Braun Channel 1 Infusion Set, it is 0.07(10ml/h), 0.02(50 ml/h), and 0.03 (100ml/h). Then on Channel 2 0.07 (10ml/hour), 0.02(50 ml/hour), and 0.1(100ml/hour). The conclusion of this study is that the use of 2 different infusion sets greatly affects the readings, other than that other factors can also affect the readings including the position of the hose and the placement of sensors on each channel. From the manufacture of this tool, it is expected that users can be more efficient in using a 2-channel Infusion Device analyzer which can be run at the same time.
Temperature Distribution Monitoring on Blood Bank Chamber Using Android Application on Mobile Phone Dianti Mayasari; Syaifuddin Syaifudin; Dyah Titisari; Triwiyanto Triwiyanto
Jurnal Teknokes Vol 16 No 1 (2023): March
Publisher : Jurusan Teknik Elektromedik, POLTEKKES KEMENKES Surabaya, Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/teknokes.v16i1.506

Abstract

Blood cold chain is a mandatory requirement of blood donation procedures to protect blood from bacterial contamination and to extend the shelf life of blood. Blood bank as a storage medium for blood bags requires a temperature between 2℃-6℃ on average. However, in general, blood banks only have 1 cold temperature distribution point, which is feared that the spread of cold temperatures in the compartment will be different at each point. For this reason, the researcher intended to design a blood bank temperature distribution monitoring device consisting of 7 measurement points. In this case, temperature sensor readings at each point are displayed wirelessly to smartphone devices using the Blynk platform and are also on a 3.5-inch TFT screen. The measurement data were then stored on the SD card memory so that the level of temperature fluctuations in the blood bank compartment can be analyzed during use. The module was also equipped with an alarm warning on the module and the Blynk application if the temperature is out of the normal temperature range (2℃-6℃). Before being used for measuring temperature distribution, the device made was compared with the standard Fluke DPM4 tester, in which the highest error obtained was 2.08% at T1, 1.58% at T2, -2.73% at T3, 1,61% at T4, -1.07% at T5, -0.06% at T6, and -2.32% at T7. After being compared with standard equipment, the device was used to measure the temperature spread in the Kirsch brand blood bank and the average temperature obtained was 3.56℃ at T1, 3.58℃ at T2, 3.73℃ at T3, 3.57℃ at T4, 3.67℃ at T5, 3.63℃ at T6, and 3.72℃ at T7. Based on the analysis results, the blood bank monitoring device can be used to measure the temperature spread in the blood bank compartment at 7 measurement points. Furthermore, temperature readings can be monitored wirelessly and remotely. It is hoped that this research can further help laboratory personnel at the Blood Transfusion Unit to monitor and evaluate the level of temperature spread in the blood bank compartment and prevent early damage to blood components.
Comparative Analysis of Water and Oil Media on Temperature Stability in PID Control-Based Digital Thermometer Calibrator Mochammad Sofyan; Syaifudin Syaifudin; Andjar Pudji; Anggara Trisna Nugraha; Bedjo Utomo
Indonesian Journal of Electronics, Electromedical Engineering, and Medical Informatics Vol 5 No 2 (2023): May
Publisher : Department of electromedical engineering, Health Polytechnic of Surabaya, Ministry of Health Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/ijeeemi.v5i2.274

Abstract

Digital thermometers are measuring instruments needed to perform temperature measurement actions and must be calibrated periodically according to standard measurement methods. The purpose of developing this tool is to add PID control to the calibration media where PID control aims to regulate the stability of the temperature setting to be achieved. This is achieved by studying and evaluating the effect of temperature stability on the heater and LM35DZ temperature sensor. This research method uses the Arduino Nanosystem for data processing and PID system control. The LM35DZ temperature sensor on the heater is regulated by a 2 Channel SSR module using a PID system then the temperature generated by the heater will be read by the LM35DZ and displayed on the LCD. The results of this study, digital thermometer calibrator measurements have been successfully carried out by comparing 3 digital thermometers with different brands, namely Omron 343F, Omron 245, and ThermoOne. The difference in error values in oil media is 3-4% and in water media 2-4% with the value of time stability in water media for 3-3.3 minutes and in oil media for 1-2 hours. With this comparison of calibration media, it is hoped that it can help in measuring temperature with better and more effective results. find methods, results, conclusions.
Bedside Monitor Based on Personal Computer Using STM32F7 Microcontroller Ingga Ariestya Riswandhani; Priyambada Cahya Nugraha; Syaifudin Syaifudin
Jurnal Teknokes Vol 16 No 2 (2023): June
Publisher : Jurusan Teknik Elektromedik, POLTEKKES KEMENKES Surabaya, Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/teknokes.v16i2.566

Abstract

A vital sign examination is one of the important indications used to establish the diagnosis of a disease and is useful for determining the medical treatment plan needed for the patient. An electrocardiograph (ECG) is a parameter in medical equipment used in the process of measuring the electrical activity of the heart muscle by measuring biopotential differences from the body surface. In 2016, cardiovascular disease was the number-one cause of death in the world. This happens because the detection of cardiovascular disease is often late, so a monitoring tool is needed that can monitor the patient's condition quickly and efficiently. The purpose of this research is to create a tool that is used to facilitate the monitoring of patient conditions. The implication of this research is that in the many cases where the signal produced is not perfect and there is still a lot of noise, this can be overcome by using the STM32F7 microcontroller, which has a 16-bit resolution, so that the resulting signal will be better, smoother, and have less noise. produced is very small. The method used in this study was to use a phantom ECG as a comparison and to use five respondents whose BPM values were to be compared with another comparison using a pulse oximeter. The design of this tool uses an ECG analog circuit that is placed on the patient's lead II leads to detect the patient's electrocardiograph signal. Data processing will be done using the STM32F7 microcontroller, and the results of the data processing will be sent to the PC using Visual Basic. The results showed that the BPM error value using a phantom ECG was 2.5%. While the smallest error value is 0,83%. In BPM measurements using 5 respondents, the largest error value was 0.9% and the smallest error value was 0%. The results of these tests indicate that this module can be used to monitor the value of each parameter in accordance with the plan.
PID Temperature Control on Blood Warmer Equipped with Patient Temperature and Blood Temperature Clarissa Grace Santoso; Torib Hamzah; Syaifudin Syaifudin; Muhammad Umer Farooq Mujahid
Indonesian Journal of Electronics, Electromedical Engineering, and Medical Informatics Vol 5 No 3 (2023): August
Publisher : Department of electromedical engineering, Health Polytechnic of Surabaya, Ministry of Health Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/ijeeemi.v5i3.286

Abstract

Body temperature in humans varies greatly depending on the location where the reading is taken. Normal core body temperature in humans is maintained by the hypothalamus and usually ranges from 36.5°C to 37.5°C. One of the causes of failure in the blood transfusion process can cause death in humans, one of the factors is the blood temperature that is too high or too low during the blood transfusion process can cause the blood to become frozen or damaged, therefore the purpose of this tool is to lower the blood temperature admission to the patient can be achieved so that there is no reduction in temperature or decrease in temperature and so that the blood is not allowed to get too hot because it can cause damage to red blood cells. This study uses the DS18B20 Sensor to control the heater with PID and Fuzzy controls, the MLX90614 Sensor to set the temperature according to the patient's body temperature and the Optocoupler Sensor as an indicator when fluids run out. When using the PID control with Kp = 4, Ki = 1, and Kd = 4, a faster response time is obtained and there is an overshoot with the highest error value of 0.77 and an average error value of 0.02. The results of the study are displayed on the TFT Nextion. From the results of the research above, it can be concluded that using PID control the response time is faster, but there are drawbacks to high overshoot.
Accuracy of Infrared Photodiode Sensors at The Flowrate Measurement in Infusion Device Analyzer with 2 Channel TFT Display Wafiq Nur Azizah; Triana Rahmawati; Syaifudin Syaifudin
Indonesian Journal of Electronics, Electromedical Engineering, and Medical Informatics Vol 5 No 3 (2023): August
Publisher : Department of electromedical engineering, Health Polytechnic of Surabaya, Ministry of Health Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/ijeeemi.v5i3.306

Abstract

The use of infusion is crucial for patient healing. Infusion refers to a fluid that consists of drugs, nutrients, and hydration delivered continuously into the patient's bloodstream over a specific period. One of the types of infusion devices is the infusion pump and syringe pump. These devices play a vital role in accurately and precisely controlling the volume or flow rate of fluids. However, continuous usage of these devices can sometimes result in inaccurate measurements, which can affect their overall accuracy. The accuracy of these devices is crucial for proper dosage administration to patients, particularly in critical situations. Therefore, it is necessary to periodically calibrate healthcare devices, at least once a year, as specified in Ministry of Health Regulation No. 54 of 2015. Calibration is an activity performed to determine the true value of a device. The objective of this study is to develop an Infusion Device Analyzer (IDA) with a TFT LCD display that showcases graphical representations of flow rate parameters. By analyzing the calculation of flow rate values using Infrared Photodiode sensors, the stability of the flow rate graph can be observed on a 7-inch TFT LCD display. The measurement involved the use of two different brands of syringe pumps and two different brands of infusion pumps. The results were presented in real-time on the 7-inch TFT LCD display, both in graphical and numerical formats. Additionally, the data was transmitted via Bluetooth to a PC, allowing the graph to be simultaneously displayed in a Delphi program.The measurement results revealed performance errors when using the Terumo Syringe Pump with Terumo syringes in Channel 1, with values of 0.45% (10 ml/h), 0.72% (50 ml/h), and 0.40% (100 ml/h). In Channel 2, the errors were 0.32% (10 ml/h), 0.40% (50 ml/h), and 0.32% (100 ml/h). When using the B-Braun Syringe Pump with B-Braun syringes, Channel 1 exhibited errors of 0.45% (10 ml/h), 0.7% (50 ml/h), and 0.85% (100 ml/h), while Channel 2 had errors of 0.8% (10 ml/h), 0.3% (50 ml/h), and 1% (100 ml/h). In the case of the Terumo Infusion Pump with Terumo Infusion Sets, Channel 1 showed errors of 0.4% (10 ml/h), 0.5% (50 ml/h), and 0.45% (100 ml/h), and Channel 2 exhibited errors of 0.32% (10 ml/h), 0.4% (50 ml/h), and 0.72% (100 ml/h). Lastly, when using the B-Braun Infusion Pump with B-Braun Infusion Sets, Channel 1 had errors of 0.72% (10 ml/h), 1% (50 ml/h), and 1,2% (100 ml/h), while Channel 2 displayed errors of 0.8% (10 ml/h), 0.72% (50 ml/h), and 0,4% (100 ml/h).INDEX TERMS Infrared Photodiode Sensor, Calibration, Real Time, Flow Rate.
The Use of Two DPM Modes Equipped with a Thermohygrometer and Pressure Selection for the Implementation of Community Service Syaifudin Syaifudin; Ketut Dyah Kusumawati; Tribowo Indrato
Frontiers in Community Service and Empowerment Vol. 1 No. 1 (2022): March
Publisher : Forum Ilmiah Teknologi dan Ilmu Kesehatan (FORITIKES)

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/ficse.v1i1.5

Abstract

Calibration is a technical activity consisting of determination, determination of one or more properties and characteristics of a product, process,s or service in accordance with a specific procedure that has been determined. The purpose of calibration is to ensure that the measurement results comply with national and international standards. Digital Pressure Meter (DPM) Two Modes equipped with Thermohygrometer and Pressure Selection is a device used to determine positive pressure and negative pressure (vacuum). Using the MPX 5050GP sensor as a positive pressure sensor. maximum pressure requirement is 300 mmHg. Using a DHT 22 sensor for temperature humidity. This device is also equipped with a leak test timer. The display used in this module is a 2.4-inch TFT Nextion LCD. After carrying out the measurement process to the mercury sphygmomanometer 6 times, the smallest result was 0 mmHg at 0 mmHg pressure and the largest was 298.0 mmHg at 300mmHg pressure. Based on the temperature humidity data collection carried out in the room obtained 31̊C at temperature and 87% at humidity.
A Fuzzy Logic Approach for Temperature Stability in Infant Incubator Systems Syaifudin; Abd. Kholiq; Anita Miftahul Maghfiroh
Engineering Science Letter Vol. 4 No. 03 (2025): Engineering Science Letter
Publisher : The Indonesian Institute of Science and Technology Research

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.56741/IISTR.esl.001451

Abstract

An infant incubator is a critical life-support device that provides thermal regulation for premature or low-birth-weight infants who are unable to maintain stable body temperature. Precise temperature control is essential, as instability and prolonged transient responses can increase health risks. Conventional on-off control methods commonly used in basic incubator systems often result in slower stabilization and higher temperature error. Therefore, this study aims to design and implement a baby incubator temperature control system using a fuzzy logic controller (FLC) integrated with a DS18B20 temperature sensor to improve thermal stability. The proposed system was implemented on a physical incubator prototype and evaluated experimentally. System performance was assessed based on dynamic response characteristics, including rise time, peak overshoot, and settling time. Experiments were conducted using three temperature setpoints: 32°C, 35°C, and 36°C. To ensure measurement accuracy, system performance was validated using an Incu Analyzer as a reference device. The experimental results show that the fuzzy logic-based control system achieved a steady-state temperature error of approximately 1% across all setpoints. The maximum observed settling time after peak overshoot was 100 seconds, indicating faster and more stable temperature regulation compared with conventional on-off control methods reported in previous studies. These results demonstrate that fuzzy logic control is effective in handling nonlinear thermal dynamics and improving temperature stability in infant incubator systems. This study focuses on technical performance evaluation; therefore, further investigations related to safety assessment and regulatory compliance are required before clinical implementation. Nevertheless, the proposed system shows strong potential as an intelligent temperature control approach for the development of neonatal incubator technology.
Development of an IoT-Based Anthropometric System Employing K-Means Clustering for Stunting Detection and Spatial Mapping in Toddlers Syaifudin Syaifudin; Endro Yulianto; Miftakhul Huda Wildany
Jurnal Teknokes Vol. 19 No. 2 (2026): June
Publisher : Jurusan Teknik Elektromedik, Politeknik Kesehatan Kemenkes Surabaya, Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/jteknokes.v19i2.149

Abstract

Stunting remains a major public health challenge in Indonesia, affecting children’s physical growth and cognitive development due to inaccurate and delayed monitoring in community health centers. This study aims to develop an Internet of Things (IoT)-based anthropometric measurement and regional stunting mapping system that provides real-time, automated, and spatially contextualized data analysis. The novelty of this research lies in integrating IoT sensor networks with machine learning–based K-Means clustering and statistical validation through the Sum of Squared Error (SSE) method, supported by an automated email alert for high-risk areas. Unlike previous studies that focus solely on anthropometric measurement or standalone IoT monitoring, this study integrates real-time IoT-based data acquisition with K-Means clustering for regional stunting mapping and automated alert generation. The system employs an HC-SR04 ultrasonic sensor, an MPU-6050 gyroscope, and an ESP32 microcontroller for data acquisition and transmission, followed by clustering analysis to categorize stunting prevalence into five levels. Experimental results show high measurement accuracy (mean error of 1.24%) and optimal clustering compactness (SSE = 1.72 × 10³ at k = 5), effectively identifying regions with very high prevalence and visualizing them through a web-based dashboard. Although the study is limited by the use of secondary datasets and pilot- scale validation, the findings demonstrate that the proposed IoT-based framework can enhance data-driven public health decision-making. This innovation aligns with Indonesia’s national stunting reduction strategy and supports Sustainable Development Goal (SDG) 3 Good Health and Well-being, contributing to the digital transformation of early childhood health monitoring.