Analysis of Temperature Distribution in Blood Banks Through Storage of Measurement Results with IoT Monitoring in the Blood Donor Unit of PMI Surabaya Farisy Azis Satria Wardhana; Anita Miftahul Maghfiroh; Dyah Titisari; Sumber Sumber; Vugar Abdullayev
Indonesian Journal of Electronics, Electromedical Engineering, and Medical Informatics Vol 5 No 2 (2023): May
Publisher : Department of electromedical engineering, Health Polytechnic of Surabaya, Ministry of Health Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/ijeeemi.v5i2.278

Abstract

Temperature or temperature is an indicator of the degree of heat of an object. Cold chain or cold chain is a supply chain system that considers the temperature level in the process. Cold chain to keep frozen or chilled products in an environment with a certain temperature during production, storage, transportation, processing and sales. This is intended to maintain product quality. The purpose of this study was to determine the temperature distribution in the Blood bank at UDD PMI Surabaya City which was used for storage of blood products. By using the ESP32 system and the DS18B20 temperature sensor which will then be monitored via IoT, it will make it easier for users to monitor. The results of these measurements will be stored in a micro SD card for analysis. The data is processed by Non-Parametric Test resulting in an interpretation that the temperature of each shelf is different due to the difference in the location of the sensor placement. The temperature difference is also influenced by the pattern of use and the process of heat transfer from the bottom to the top of the shelf. This research was considered successful with the result of the highest temperature distribution being 3°C and the lowest being 2°C. The location of these racks can be useful in determining day-to-day monitoring measuring points. This value has met the standard for storage of blood products, which is in the range of 2°C-6°C.
Guidance And Assistance In Repair Of Centrifuge Laboratory Equipment Based On Standards Of Centrifuge Motor Speed Data In Turi Health Centre, Lamongan Regency, East Java I Dewa Gede Hari Wisana; Her Gumiwang Ariswati; Sumber Sumber; Levana Forra Wakidi
Frontiers in Community Service and Empowerment Vol. 1 No. 1 (2022): March
Publisher : Forum Ilmiah Teknologi dan Ilmu Kesehatan (FORITIKES)

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/ficse.v1i1.3

Abstract

One of the equipment in the Turi Health Center laboratory is a Centrifuge. The problem at the Turi Health Center with the centrifuge is that this centrifuge has a large usage load of 8-10 hours per day, without having electromedical personnel. In this Centrifuge, it is often found that the cleanliness of the tools is not good, the placement of the tools is inadequate, the electrical installation network is not standard (Grounding) and the maintenance and maintenance is lacking, there is no regular calibration so there is no certainty of the accuracy of the results obtained. As a result of the impact obtained is a tool that is not suitable for use, the tool is quickly damaged which can harm the diagnosis of patients. Community Service Activities with the implementation team from Poltekkes Kemenkes Surabaya working together as partners in this activity aims to carry out repairs according to Centrifuge standards and provide knowledge and skills about simple maintenance and repairs on Centrifuges Correctly. The first stage in providing solutions to problems faced by the Turi Health Center is to conduct a study through Focus Group Discuss (FGD). The second stage is compiling a user manual. The third stage is the implementation of mentoring and counseling on centrifuge repair using a maintenance and repair manual. And the fourth stage is the implementation of centrifuge repair by providing opportunities for laboratory personnel, Medical and non-medical personnel to be able to independently carry out simple repairs and to know important indicators in the event of damage. The output target of the community service carried out is a simple repair guide book for centrifuge equipment so that later laboratory equipment at the Turi Health Center, especially the Centrifuge, can work according to the Centrifuge Motor Speed Standard.
Internal Calibration For Centifuge Using A Digital Tachometer as a Synergistic Collaboration Between the Health Polytechnic of the Ministry of Health Surabaya and the Turi Lamongan Health Center Her Gumiwang Ariswati; I Dewa Hari Wisana; Sumber Sumber; Levana Fora Wakidi
Frontiers in Community Service and Empowerment Vol. 1 No. 1 (2022): March
Publisher : Forum Ilmiah Teknologi dan Ilmu Kesehatan (FORITIKES)

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/ficse.v1i1.4

Abstract

Centrifuge is a laboratory tool used for the process of separating particles of a solution between the filtrate and the substrate through a high-speed centripetal process. The increasing use of Centrifuges at the Turi Lamongan Health Center causes the equipment to be vulnerable to damage. In addition, the lack of handling of these tools is due to limited medical analysts and the absence of electromedical personnel as well as insufficient funds, so that the impact on the health service system at the puskesmas is disrupted. Therefore, it is very important to create a community service program that is carried out in the form of a synergistic collaboration between the Health Polytechnic of the Ministry of Health Surabaya and the Turi Lamongan Health Center which is oriented towards community independence through counseling and assistance, especially in the maintenance of the Centrifuge laboratory. The purpose of this activity is maintenance by testing and calibrating/verifying the motor speed and time on the centrifuge in accordance with the working guidelines. The methodological concept in this activity is to carry out counseling and assistance activities in the maintenance and verification of centrifuge laboratory equipment using a digital tachometer. This activity is an application of previous research on centrifuges and tachometers. The tools used are digital tachometer, savety electrical analyzer, avometer, thermohygrometer, stopwatch and tool set. The materials used are several spare parts, compressors, cleaning tools and cleaning fluids as well as antiseptic fluids. The results obtained from these activities are: 1). The technical specifications of the 4 centrifuge devices are recorded; 2). Identification of damaged centrifuges and those requiring maintenance, namely 2 heavily damaged centrifuges and 2 centrifuges that are still suitable for use; 3). Maintenance of 4 centrifuges; 4). Verification/calibration of 2 centrifuges, namely centrifugal 1 with a relative error of 2.73% in the selection of a motor rotation speed of 2500 rpm is still within the allowable limits (≤10%); for testing the time of 300 seconds, the relative error of 20.57% exceeds the allowable limit, while in centrifuge 2 the results of the relative error in 4 selections of motor rotation speed (rpm) are below 10%, namely 5.92%, respectively 3.82. %, 3.65%, 3.47%. As for the time test, the result of 5.56% is still below the maximum allowable error of 10%. From these results, centrifuge 2 is still in good condition; 5) Implementation of mentoring and counseling for 1 (one) medical laboratory technician (analyst) in performing centrifuge equipment maintenance. The results of this activity can facilitate services at the Turi Public Health Center.
Exploration of digital filters on cardiac monitor devices equipped with non-invasive blood pressure (NIBP) Nugraha, Priyambada C.; Sumber, Sumber; Muzachim, Zuva; Rabani, Rifqi; Alhaq, Elmira Rofida; Triwiyanto, Triwiyanto; Abdullayev, Vugar
Indonesian Journal of Electronics, Electromedical Engineering, and Medical Informatics Vol. 6 No. 4 (2024): November
Publisher : Jurusan Teknik Elektromedik, Politeknik Kesehatan Kemenkes Surabaya, Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/edd73780

Abstract

Heart disease is a leading cause of global mortality, making accurate monitoring essential for early detection and prevention of complications. Although heart monitoring technology has advanced, there are still limitations in precisely detecting early symptoms. This study aims to develop a Cardiac Monitor device capable of monitoring patients with heart conditions through three main parameters: electrocardiogram (ECG), phonocardiogram (PCG), and non-invasive blood pressure measurement (NIBP). The system designed in this research integrates digital filters, namely Butterworth (order 2, 4, 8) and Kalman, to enhance the quality of ECG and PCG signals. Additionally, the oscillometric method in non-invasive blood pressure measurement (NIBP) is used as a comparison for blood pressure estimation by analyzing the correlation between the R peak on the ECG signal, pulse transit time (PTT), and the first and second heart sounds (S1, S2) on the PCG signal. Blood pressure estimation is performed using algorithmic calculations to determine the accuracy of the design module in measuring systolic and diastolic pressure. The results indicate that the 8th-order Butterworth filter is more effective in reducing noise in ECG and PCG signals compared to the Kalman filter. The study also finds a significant correlation between the R peak on the ECG and the first heart sound on the PCG. The design module’s blood pressure measurement errors compared to algorithmic estimates are 4.54 ± 4.94 mmHg for systolic pressure and 6.57 ± 3.83 mmHg for diastolic pressure, which are close to the AAMI standard of 5 ± 8 mmHg. These findings highlight the great potential of the developed Cardiac Monitor device in improving early diagnosis accuracy and heart disease management.
Enhancing Laboratory Efficiency: A Community Service Program for Centrifuge Maintenance at Randegansari Husada Hospital Setioningsih, Endang Dian; Sumber; Setiawan, Singgih Yudha; Lusiana
Frontiers in Community Service and Empowerment Vol. 4 No. 1 (2025): March
Publisher : Forum Ilmiah Teknologi dan Ilmu Kesehatan (FORITIKES)

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/ficse.v4i1.90

Abstract

The community service program conducted by Poltekkes Kemenkes Surabaya aimed to address the problems associated with equipment maintenance at the Randegansari Husada Hospital, focusing on its centrifuge equipment in the laboratory. The program included counseling and provided practical assistance to the hospital staff, preparing them with the knowledge and skills for the proper maintenance of equipment. Activities covered recording the specification data of the centrifuge units, damage detection, performance improvement using a preventive maintenance program, and performance testing using a digital tachometer. The calibration process also proved a centrifuge relative error of 1.37% at 3000 RPM within the acceptable error limit of 10%. Verification in time, at 300 seconds, resulted in a relative error of 4.89%, also within acceptable limits. Such efforts go a step further to ensure that the centrifuge works well in contributing toward better diagnostic accuracies. The program took a portion of this staffing challenge when it embarked on the training of a medical laboratory technician, insisting that improved and sustained equipment reliability meant improved health service delivery. These would then increase at the hospital to limit malfunctioning of equipment for overall quality care
Motor Rotation Control System on Syringe Pump Using PID and IoT-Based Flow Rate Monitoring Tan, Ratna; Sumber, Sumber; Triwiyanto, Triwiyanto
Jurnal Teknokes Vol. 17 No. 4 (2024): December
Publisher : Jurusan Teknik Elektromedik, Politeknik Kesehatan Kemenkes Surabaya, Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar

Abstract

Syringe pump is a life support device that functions to provide drugs in the form of fluids through the vein to the patient's body in a measurable and consistent within a period of time. This research is designed to make the syringe pump more stable, equipped with PCA (Patient Controlled Analgesia) mode, and IoT-based flowrate monitoring. The contribution made by the author is to improve the working system of the syringe pump to be more stable which is equipped with a PCA mode that functions to inject liquid according to the patient's needs, and facilitate remote flowrate monitoring using IoT. In this research, the author uses the PID method to improve the stability of the motor according to the flowrate setting. From the test results that have been done, the flowrate stability of the PID system is quite good in all settings. The resulting rise time is good enough to reach the setpoint, the average occurs in seconds 1-5, the fastest rise time occurs at a flowrate setting of 5 ml / h, which is 1 second and the longest rise time occurs at a flowrate setting of 20 ml / h, which is 5 seconds. The highest overshoot occurs at a flowrate setting of 50 ml/hour with a value of 1.98 ml/hour and the lowest occurs at a flowrate of 5 ml/hour with 0.44 ml/hour. The highest and lowest stability occurs at a flowrate setting of 10 ml/hour, which is 1.5% and 6.5%. In PCA mode, the largest error obtained is around 1% in all settings. IoT-based flowrate monitoring is good, the value displayed on the LCD in real time is sent and displayed on the web. Based on the results of the data that has been obtained, the PID system is still not optimal in controlling the stability of the flowrate so that in the future it is necessary to find a suitable and balanced combination of Kp, Ki, and Kd values.     
System Motor Gearbox Control For Increase Accuracy Angle On CPM Elbow Tool PID Method Kurniawan, Desi Dwi; Ariswati, Her Gumiwang; Lamidi; Sumber
Jurnal Teknokes Vol. 18 No. 1 (2025): March
Publisher : Jurusan Teknik Elektromedik, Politeknik Kesehatan Kemenkes Surabaya, Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar

Abstract

Continuous Passive Motion (CPM) is a rehabilitation tool designed to help joint recovery after surgery by passively moving the joint within a certain range of motion. Previous studies have used the ATmega8535 microcontroller with a PWM system. In this study, a control system was developed using Proportional-Integral-Derivative (PID) to improve the angular accuracy of the CPM tool specifically for the elbow joint. The implementation of PID aims to control the gearbox motor to achieve angles of 45°, 90° and 130° which drive the CPM tool so that it can operate with high accuracy and adjust the angle according to the desired therapy settings. This tool using Encoder sensor to detect corner movement joints in real-time. The Arduino Uno microcontroller controls the motors based on feedback from the sensors and optimizes movement through PID control . Research results show that PID system can increase accuracy corner with a low average error . In addition , the system is also equipped with with measurement current using the ACS712 sensor for ensure stability Power during operation . Result of study This can read 45°, 90° and 130° angles with a good level of accuracy , besides it is also capable operated with speed of 40°/ min , 90°/ min , 150°/ min , 210°/ min and 270°/ min . In the measurement readings on the compared devices with Goniometer there is difference error at 45° angle is 5°, 90° angle is 1° and 130° angle is 5°. for measurement current on the device the obtained stable value​ 2.5mA.Research This succeed reach optimization control angle on the CPM tool with level good accuracy , so that can support the rehabilitation process patient with more effective and efficient .
Integrated Digital Sphygmomanometer for Simultaneous Blood Pressure and Body Temperature Monitoring Aulia, Farahun Nisa; Pudji, Andjar; Sumber, Sumber
Jurnal Teknokes Vol. 17 No. 2 (2024): June
Publisher : Jurusan Teknik Elektromedik, Politeknik Kesehatan Kemenkes Surabaya, Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar

Abstract

Hypertension is a condition when the blood pressure against the artery walls is too high. High blood pressure can occur because of the patient's temperature, because the higher the patient's body temperature the higher the blood pressure. Therefore, body temperature must be detected before measuring blood pressure in patients. These Digital Tension and Body Temperature parameters are usually still used separately. Because of that, we had the idea to make these two parameters into one unit to facilitate health monitoring. The design of this tool uses the MPX5050GP sensor as a blood pressure detector, the MLX90614 sensor as a body temperature detector and the two sensors are connected directly to the Arduino UNO microcontroller to be processed and later displayed on the Nextion LCD later. The measurement results with the MPX5050GP and MLX9061 showed that the largest systolic error was 2.23% and the smallest was 0.53%. The biggest diastole error was 4.69% and the smallest was 1.79%. The biggest body temperature error is 1.65% and the smallest is 0.45%.
Wireless Blood Pressure Monitor with Android Integration: Tracking Systolic and Diastolic Parameters Sintiyah, Elisa Ayu; Sumber, Sumber; Yudha Setiawan, Singgih
Jurnal Teknokes Vol. 17 No. 2 (2024): June
Publisher : Jurusan Teknik Elektromedik, Politeknik Kesehatan Kemenkes Surabaya, Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar

Abstract

Blood pressure measurement plays a crucial role in detecting underlying diseases in the human body. It enables the identification of conditions like heart failure, kidney failure, liver damage, and stroke, underscoring the importance of regular measurements. To facilitate independent and routine blood pressure monitoring, the development of an automatic blood pressure measuring device is essential. This research aims to design and fabricate a digital sphygmomanometer that can transmit measurements to a smartphone through the Blynk application. The blood pressure measurement is conducted using the MPX5050GP pressure sensor as the pressure detector. The device is programmed using the Esp32 microcontroller and incorporates an LCD screen to display the measurement results. The study involved measuring six participants, with each individual's blood pressure recorded ten times. The obtained measurements were then compared to those of the Omron HEM-7120 digital sphygmomanometer. The results revealed a discrepancy of ±9 mmHg in systolic values and ±7 mmHg in diastolic values between the two devices. The smallest systolic error observed was 0.4%, while the largest error reached 3%. Similarly, the smallest diastolic error was 2%, with the largest error recorded at 4.8%. The measurement errors, particularly in diastolic pressure, were influenced by the participants' fatigue, as the repeated measurements on the same arm led to slight arm movements during the process. The study demonstrated the successful transmission of measurement results to a smartphone, affirming the efficacy of the Blynk application. Additionally, the MPX5050GP sensor proved effective indetecting blood pressure. These findings highlight the potential of the developed digital sphygmomanometer as a reliable tool for blood pressure monitoring, promoting self-care and early detection of health issues.
Digital Sphygmomanometer Detects Systole Diastolic Display Aulia, Farahun Nisa; Pudji, Andjar; Sumber, Sumber; Ullah, Naqeeb
Jurnal Teknokes Vol. 16 No. 4 (2023): December
Publisher : Jurusan Teknik Elektromedik, Politeknik Kesehatan Kemenkes Surabaya, Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar

Abstract

Hypertension. characterized by elevated blood pressure against artery walls. can be influenced by a patient's body temperature. Therefore. detecting body temperature before measuring blood pressure is essential for accurate assessment. Currently. Digital Tension and Body Temperature parameters are typically evaluated separately. To address this. we propose a novel approach to combine these parameters into a single unit. enhancing health monitoring. Utilizing MPX5050GP for blood pressure and MLX90614 for body temperature detection. Both sensors are directly connected to the Arduino UNO microcontroller. enabling seamless data processing and display on the Nextion LCD. Experimental results demonstrate the device's effectiveness. with systolic blood pressure measurements showing a Maximum error: 2.23%. minimum error: 0.53% for systolic measurements. Diastolic measurements have with a remarkable maximum error of only 4.69% and a minimal error of 1.79%. Additionally. the body temperature measurements exhibited a Achieved exceptional precision with errors as low as 0.45% and a maximum of 1.65%. Successfully completed. this design facilitates simultaneous measurement of two vital parameters. Its potential to streamline health monitoring could significantly impact hypertension management and other related conditions. Further validation and implementation in clinical settings are anticipated to enhance its utility and benefits.
Co-Authors A. Senthil Kumar Abd . Kholiq Abdul Kholiq Abdullayev, Vugar Alhaq, Elmira Rofida Amrinsani, Farid Andjar Pudji Anita Miftahul Maghfiroh Aulia, Farahun Nisa Bambang Guruh I Bambang Guruh Irianto Bima Maulana Raharjo Caesar Febri Nugroho Candra Prastyadi Chandra Bimantara Putra Decoriza Kurnia Abadi Dimas Adi tya Dwi K, Rizky Dwi S, Frendi Agung Dyah Titisari, Dyah Elmira Rofida Al Haq Endro Yulianto Farisy Azis Satria Wardhana Febrianti, Fani Forra Wakidi, Levana Frendi Agung Dwi Saputra Gadang Hendra Prabowo Hamzah, Torib Hanna Firdaus Her Gumiwang Her Gumiwang Ariswati I Dewa Gede Hari Wisana Jing Lu Kurniawan, Desi Dwi Lamidi, Lamidi Levana Fora Wakidi Liliek Soetjiatie Lusiana Luthfiyah, Sari Mak’ruf, Muhammad Ridha Mansour Asghari Muhammad Ridha Mak'ruf Muhammad Syifaul Linnas Muzachim, Zuva Muzni Zahar Nazila Ragimova Nisa'ul Sholihah Nugraha, Priyambada C. Nurul Al Istigho Farola Prastawa Assalim Tetra Putra Priyambada Cahya Nugraha Putra, Chandra Bimantara Rabani, Rifqi Ragimova, Nazila Rangga Adi Firmansyah Rathod, Yagnik RATNA TAN Roichatun Nashichah Sella Octa Ardila Setiawan, Singgih Yudha Setioningsih, Endang Dian Sintiyah, Elisa Ayu Syafiq Naufal Syayakti Syaifudin Syaifudin , Syaifudin Syevana Dita Musvika Tetra Putra, Moch Prastawa Assalim Tetrik Fa’altin Torib Hamzah Torib Hamzah Triwiyanto Triwiyanto Ullah, Naqeeb Utomo, Bedjo Vijay Anant Athavale Vugar Abdullayev Vugar Abdullayev Wahyu Caesarendra Wahyu Caesarendra Wahyu Ramadhan Putra Wardhana, Farisy Azis Satria Zahar, Muzni